Show simple item record

contributor authorTree, Michael
contributor authorWhite, Jason
contributor authorMidha, Prem
contributor authorKiblinger, Samantha
contributor authorYoganathan, Ajit
date accessioned2017-05-09T01:31:39Z
date available2017-05-09T01:31:39Z
date issued2016
identifier issn1932-6181
identifier othermed_010_01_011001.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/161979
description abstractThe CardioMEMS heart failure (HF) system was tested for cardiac output (CO) measurement accuracy using an in vitro mock circulatory system. A software algorithm calculates CO based on analysis of the pressure waveform as measured from the pulmonary artery, where the CardioMEMS system resides. Calculated CO was compared to that from reference flow probe in the circulatory system model. CO measurements were compared over a clinically relevant range of stroke volumes and heart rates with normal, pulmonary hypertension (PH), decompensated left heart failure (DLHF), and combined DHLF + PH hemodynamic conditions. The CardioMEMS CO exhibited minimal fixed and proportional bias.
publisherThe American Society of Mechanical Engineers (ASME)
titleValidation of Cardiac Output as Reported by a Permanently Implanted Wireless Sensor
typeJournal Paper
journal volume10
journal issue1
journal titleJournal of Medical Devices
identifier doi10.1115/1.4031799
journal fristpage11001
journal lastpage11001
identifier eissn1932-619X
treeJournal of Medical Devices:;2016:;volume( 010 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record