Improving Single Molecule Experiments With Feedback Control of Optical TrapsSource: Journal of Dynamic Systems, Measurement, and Control:;2013:;volume( 135 ):;issue: 002::page 21001Author:Cole, D. G.
DOI: 10.1115/1.4007237Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: This article explores various types of feedback control—position feedback, which was shown to be equivalent to force feedback, rate feedback, and integral feedback—for the purpose of improving instrument performance for singlemolecule experiments. The ability of each of each types of feedback to lower the measurement signaltonoise ratio (SNR) is evaluated and compared to the openloop case. While position feedback does not result in any improvement in the SNR, the cases of rate feedback and integral feedback both resulted in improvements in the measurement's SNR. Rate feedback is shown to effectively “cool†the beads held in the optical trap, thereby limiting the effect that Brownian disturbances have on the beads’ motion. Integral feedback is shown to improve the SNR of the measured signal of interest and is robust and easy to implement. It is also shown that integral feedback acts as an exogenous force estimator. Ultimately, feedback does not provide better resolution as measured by SNR than an openloop filtering approach can but does provide other advantages, including the ability to control other variables and to make a more robust instrument that can be easily adapted to changes in experimental conditions or the environment.
|
Show full item record
| contributor author | Cole, D. G. | |
| date accessioned | 2017-05-09T00:57:16Z | |
| date available | 2017-05-09T00:57:16Z | |
| date issued | 2013 | |
| identifier issn | 0022-0434 | |
| identifier other | ds_135_2_021001.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/151264 | |
| description abstract | This article explores various types of feedback control—position feedback, which was shown to be equivalent to force feedback, rate feedback, and integral feedback—for the purpose of improving instrument performance for singlemolecule experiments. The ability of each of each types of feedback to lower the measurement signaltonoise ratio (SNR) is evaluated and compared to the openloop case. While position feedback does not result in any improvement in the SNR, the cases of rate feedback and integral feedback both resulted in improvements in the measurement's SNR. Rate feedback is shown to effectively “cool†the beads held in the optical trap, thereby limiting the effect that Brownian disturbances have on the beads’ motion. Integral feedback is shown to improve the SNR of the measured signal of interest and is robust and easy to implement. It is also shown that integral feedback acts as an exogenous force estimator. Ultimately, feedback does not provide better resolution as measured by SNR than an openloop filtering approach can but does provide other advantages, including the ability to control other variables and to make a more robust instrument that can be easily adapted to changes in experimental conditions or the environment. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Improving Single Molecule Experiments With Feedback Control of Optical Traps | |
| type | Journal Paper | |
| journal volume | 135 | |
| journal issue | 2 | |
| journal title | Journal of Dynamic Systems, Measurement, and Control | |
| identifier doi | 10.1115/1.4007237 | |
| journal fristpage | 21001 | |
| journal lastpage | 21001 | |
| identifier eissn | 1528-9028 | |
| tree | Journal of Dynamic Systems, Measurement, and Control:;2013:;volume( 135 ):;issue: 002 | |
| contenttype | Fulltext |