    Comparison of Strain-Gage and Fiber-Optic Goniometry for Measuring Knee Kinematics During Activities of Daily Living and Exercise

    Source: Journal of Biomechanical Engineering:;2012:;volume( 134 ):;issue: 008::page 84502
    Author:
    Abeer A. Mohamed
    ,
    James Beyea
    ,
    John Landry
    ,
    Chris A. McGibbon
    ,
    Andrew Sexton
    ,
    Jennifer Baba
    DOI: 10.1115/1.4007094
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: There is increasing interest in wearable sensor technology as a tool for rehabilitation applications in community or home environments. Recent studies have focused on evaluating inertial based sensing (accelerometers, gyroscopes, etc.) that provide only indirect measures of joint motion. Measurement of joint kinematics using flexible goniometry is more direct, and still popular in laboratory environments, but has received little attention as a potential tool for wearable systems. The aim of this study was to compare two goniometric devices: a traditional strain-gauge flexible goniometer, and a fiberoptic flexible goniometer, for measuring dynamic knee flexion/extension angles during activity of daily living: chair rise, and gait; and exercise: deep knee bends, against joint angles computed from a “gold standard” Vicon motion tracking system. Six young adults were recruited to perform the above activities in the lab while wearing a goniometer on each knee, and reflective markers for motion tracking. Kinematic data were collected simultaneously from the goniometers (one on each leg) and the motion tracking system (both legs). The results indicate that both goniometers were within 2–5 degrees of the Vicon angles for gait and chair rise. For some deep knee bend trials, disagreement with Vicon angles exceeded ten degrees for both devices. We conclude that both goniometers can record ADL knee movement faithfully and accurately, but should be carefully considered when high (>120 deg) knee flexion angles are required.
    keyword(s): Knee , Motion , Strain gages , Kinematics AND Sensors ,
    • Download: (922.5Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Comparison of Strain-Gage and Fiber-Optic Goniometry for Measuring Knee Kinematics During Activities of Daily Living and Exercise

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/148226
    Collections
    • Journal of Biomechanical Engineering

    Show full item record

    contributor authorAbeer A. Mohamed
    contributor authorJames Beyea
    contributor authorJohn Landry
    contributor authorChris A. McGibbon
    contributor authorAndrew Sexton
    contributor authorJennifer Baba
    date accessioned2017-05-09T00:48:26Z
    date available2017-05-09T00:48:26Z
    date copyrightAugust, 2012
    date issued2012
    identifier issn0148-0731
    identifier otherJBENDY-29000#084502_1.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/148226
    description abstractThere is increasing interest in wearable sensor technology as a tool for rehabilitation applications in community or home environments. Recent studies have focused on evaluating inertial based sensing (accelerometers, gyroscopes, etc.) that provide only indirect measures of joint motion. Measurement of joint kinematics using flexible goniometry is more direct, and still popular in laboratory environments, but has received little attention as a potential tool for wearable systems. The aim of this study was to compare two goniometric devices: a traditional strain-gauge flexible goniometer, and a fiberoptic flexible goniometer, for measuring dynamic knee flexion/extension angles during activity of daily living: chair rise, and gait; and exercise: deep knee bends, against joint angles computed from a “gold standard” Vicon motion tracking system. Six young adults were recruited to perform the above activities in the lab while wearing a goniometer on each knee, and reflective markers for motion tracking. Kinematic data were collected simultaneously from the goniometers (one on each leg) and the motion tracking system (both legs). The results indicate that both goniometers were within 2–5 degrees of the Vicon angles for gait and chair rise. For some deep knee bend trials, disagreement with Vicon angles exceeded ten degrees for both devices. We conclude that both goniometers can record ADL knee movement faithfully and accurately, but should be carefully considered when high (>120 deg) knee flexion angles are required.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleComparison of Strain-Gage and Fiber-Optic Goniometry for Measuring Knee Kinematics During Activities of Daily Living and Exercise
    typeJournal Paper
    journal volume134
    journal issue8
    journal titleJournal of Biomechanical Engineering
    identifier doi10.1115/1.4007094
    journal fristpage84502
    identifier eissn1528-8951
    keywordsKnee
    keywordsMotion
    keywordsStrain gages
    keywordsKinematics AND Sensors
    treeJournal of Biomechanical Engineering:;2012:;volume( 134 ):;issue: 008
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian