Show simple item record

contributor authorAbeer A. Mohamed
contributor authorJames Beyea
contributor authorJohn Landry
contributor authorChris A. McGibbon
contributor authorAndrew Sexton
contributor authorJennifer Baba
date accessioned2017-05-09T00:48:26Z
date available2017-05-09T00:48:26Z
date copyrightAugust, 2012
date issued2012
identifier issn0148-0731
identifier otherJBENDY-29000#084502_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/148226
description abstractThere is increasing interest in wearable sensor technology as a tool for rehabilitation applications in community or home environments. Recent studies have focused on evaluating inertial based sensing (accelerometers, gyroscopes, etc.) that provide only indirect measures of joint motion. Measurement of joint kinematics using flexible goniometry is more direct, and still popular in laboratory environments, but has received little attention as a potential tool for wearable systems. The aim of this study was to compare two goniometric devices: a traditional strain-gauge flexible goniometer, and a fiberoptic flexible goniometer, for measuring dynamic knee flexion/extension angles during activity of daily living: chair rise, and gait; and exercise: deep knee bends, against joint angles computed from a “gold standard” Vicon motion tracking system. Six young adults were recruited to perform the above activities in the lab while wearing a goniometer on each knee, and reflective markers for motion tracking. Kinematic data were collected simultaneously from the goniometers (one on each leg) and the motion tracking system (both legs). The results indicate that both goniometers were within 2–5 degrees of the Vicon angles for gait and chair rise. For some deep knee bend trials, disagreement with Vicon angles exceeded ten degrees for both devices. We conclude that both goniometers can record ADL knee movement faithfully and accurately, but should be carefully considered when high (>120 deg) knee flexion angles are required.
publisherThe American Society of Mechanical Engineers (ASME)
titleComparison of Strain-Gage and Fiber-Optic Goniometry for Measuring Knee Kinematics During Activities of Daily Living and Exercise
typeJournal Paper
journal volume134
journal issue8
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4007094
journal fristpage84502
identifier eissn1528-8951
keywordsKnee
keywordsMotion
keywordsStrain gages
keywordsKinematics AND Sensors
treeJournal of Biomechanical Engineering:;2012:;volume( 134 ):;issue: 008
contenttypeFulltext