An MRI-Compatible Needle Manipulator Concept Based on Elastically Averaged Dielectric Elastomer Actuators for Prostate Cancer Treatment: An Accuracy and MR-Compatibility Evaluation in PhantomsSource: Journal of Medical Devices:;2009:;volume( 003 ):;issue: 003::page 31005Author:Daniel F. Kacher
,
Joseph R. Roebuck
,
Simon P. DiMaio
,
Jean-Sébastien Plante
,
Kenjiro Tadakuma
,
Steven Dubowsky
,
Lauren M. DeVita
,
Ferenc A. Jolesz
DOI: 10.1115/1.3191729Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: A parallel manipulator concept using bistable polymer actuators has been developed to perform prostate cancer biopsy, and deliver therapy within the bore of a magnetic resonance imaging (MRI) scanner. The dielectric elastomer actuators (DEAs) used in this manipulator concept are promising for MRI-compatible robotics because they do not interfere with the high magnetic fields of MRI while having good mechanical performance and being low cost. In the past, these actuators have been plagued by robustness problems when used in a continuous manner. Recent studies show that reliability significantly improves when DEAs are used in a bistable manner, as proposed here. This paper investigates the potential of the proposed manipulator concept by evaluating the positioning accuracy and MRI compatibility of a laboratory prototype, developed for clinically relevant design criteria. An analytical model of the manipulator kinematics is presented. Analytical and experimental results validate that the proposed technology can provide an accurate needle placement required to perform prostate cancer treatments. The prototype’s MRI compatibility is validated in a 3 T clinical MRI scanner. The parallel manipulator concept using bistable polymer actuators is shown to be a viable approach to perform MRI-guided needle insertions for prostate cancer biopsy and therapy.
|
Collections
Show full item record
| contributor author | Daniel F. Kacher | |
| contributor author | Joseph R. Roebuck | |
| contributor author | Simon P. DiMaio | |
| contributor author | Jean-Sébastien Plante | |
| contributor author | Kenjiro Tadakuma | |
| contributor author | Steven Dubowsky | |
| contributor author | Lauren M. DeVita | |
| contributor author | Ferenc A. Jolesz | |
| date accessioned | 2017-05-09T00:34:40Z | |
| date available | 2017-05-09T00:34:40Z | |
| date copyright | September, 2009 | |
| date issued | 2009 | |
| identifier issn | 1932-6181 | |
| identifier other | JMDOA4-28006#031005_1.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/141535 | |
| description abstract | A parallel manipulator concept using bistable polymer actuators has been developed to perform prostate cancer biopsy, and deliver therapy within the bore of a magnetic resonance imaging (MRI) scanner. The dielectric elastomer actuators (DEAs) used in this manipulator concept are promising for MRI-compatible robotics because they do not interfere with the high magnetic fields of MRI while having good mechanical performance and being low cost. In the past, these actuators have been plagued by robustness problems when used in a continuous manner. Recent studies show that reliability significantly improves when DEAs are used in a bistable manner, as proposed here. This paper investigates the potential of the proposed manipulator concept by evaluating the positioning accuracy and MRI compatibility of a laboratory prototype, developed for clinically relevant design criteria. An analytical model of the manipulator kinematics is presented. Analytical and experimental results validate that the proposed technology can provide an accurate needle placement required to perform prostate cancer treatments. The prototype’s MRI compatibility is validated in a 3 T clinical MRI scanner. The parallel manipulator concept using bistable polymer actuators is shown to be a viable approach to perform MRI-guided needle insertions for prostate cancer biopsy and therapy. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | An MRI-Compatible Needle Manipulator Concept Based on Elastically Averaged Dielectric Elastomer Actuators for Prostate Cancer Treatment: An Accuracy and MR-Compatibility Evaluation in Phantoms | |
| type | Journal Paper | |
| journal volume | 3 | |
| journal issue | 3 | |
| journal title | Journal of Medical Devices | |
| identifier doi | 10.1115/1.3191729 | |
| journal fristpage | 31005 | |
| identifier eissn | 1932-619X | |
| tree | Journal of Medical Devices:;2009:;volume( 003 ):;issue: 003 | |
| contenttype | Fulltext |