Show simple item record

dc.contributor.authorAzimi, Shahab
dc.contributor.authorArzanpour, Siamak
dc.date.accessioned2024-04-24T22:31:17Z
dc.date.available2024-04-24T22:31:17Z
dc.date.copyright2/7/2024 12:00:00 AM
dc.date.issued2024
dc.identifier.issn0148-0731
dc.identifier.otherbio_146_04_041002.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4295375
dc.description.abstractThis paper presents an innovative approach to the design optimization of valved holding chambers (VHCs), crucial devices for aerosol drug delivery. We present the design of an optimal cylindrical VHC body and introduce a novel valve based on particle impaction theory. The research combines computational simulations and physical experiments to assess the performance of various VHCs, with a special focus on the deposition patterns of medication particles within these devices. The methodology incorporates both experimental and simulation approaches to validate the reliability of the simulation. Emphasis is placed on the deposition patterns observed on the VHC walls and the classification of fine and large particles for salbutamol sulfate particles. The study reveals the superior efficacy of our valve design in separating particles compared to commercially available VHCs. In standard conditions, our valve design allows over 95% of particles under 7 μm to pass through while effectively filtering those larger than 8 μm. The optimized body design accomplishes a 60% particle mass flow fraction at the outlet and an average particle size reduction of 58.5%. When compared numerically in terms of size reduction, the optimal design outperforms the two commercially available VHCs selected. This study provides valuable insights into the optimization of VHC design, offering significant potential for improved aerosol drug delivery. Our findings demonstrate a new path forward for future studies, aiming to further optimize the design and performance of VHCs for enhanced pulmonary drug delivery.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleEnhancing Inhalation Drug Delivery: A Comparative Study and Design Optimization of a Novel Valved Holding Chamber
dc.typeJournal Paper
dc.journal.volume146
dc.journal.issue4
dc.journal.titleJournal of Biomechanical Engineering
dc.identifier.doi10.1115/1.4064436
dc.journal.fristpage41002-1
dc.journal.lastpage41002-11
dc.page11
dc.treeJournal of Biomechanical Engineering:;2024:;volume( 146 ):;issue: 004