Show simple item record

dc.contributor.authorTan, Wei Ting
dc.contributor.authorLiew, Yih Miin
dc.contributor.authorMohamed Mokhtarudin, Mohd Jamil
dc.contributor.authorPirola, Selene
dc.contributor.authorWan Ab Naim, Wan Naimah
dc.contributor.authorAmry Hashim, Shahrul
dc.contributor.authorXu, Xiao Yun
dc.contributor.authorLim, Einly
dc.date.accessioned2022-02-06T05:28:34Z
dc.date.available2022-02-06T05:28:34Z
dc.date.copyright5/3/2021 12:00:00 AM
dc.date.issued2021
dc.identifier.issn0148-0731
dc.identifier.otherbio_143_08_081005.pdf
dc.identifier.urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4278108
dc.description.abstractA computational approach is used to investigate potential risk factors for distal stent graft-induced new entry (dSINE) in aortic dissection (AD) patients. Patient-specific simulations were performed based on computed tomography images acquired from six AD patients (three dSINE and three non-dSINE) to analyze the correlation between anatomical characteristics and stress/strain distributions. Sensitivity analysis was carried out using idealized models to independently assess the effect of stent graft length, stent tortuosity and wedge apposition angle at the landing zone on key biomechanical variables. Mismatch of biomechanical properties between the stented and nonstented regions led to high stress at the distal stent graft–vessel interface in all patients, as well as shear strain in the neighboring region, which coincides with the location of tear formation. Stress was observed to increase with the increase of stent tortuosity (from 263 kPa at a tortuosity angle of 50 deg to 313 kPa at 30 deg). It was further amplified by stent graft landing at the inflection point of a curve. Malapposition of the stent graft led to an asymmetrical segment within the aorta, therefore changing the location and magnitude of the maximum von Mises stress substantially (up to +25.9% with a +25 deg change in the distal wedge apposition angle). In conclusion, stent tortuosity and wedge apposition angle serve as important risk predictors for dSINE formation in AD patients.
dc.publisherThe American Society of Mechanical Engineers (ASME)
dc.titleEffect of Vessel Tortuosity on Stress Concentration at the Distal Stent–Vessel Interface: Possible Link With New Entry Formation Through Biomechanical Simulation
dc.typeJournal Paper
dc.journal.volume143
dc.journal.issue8
dc.journal.titleJournal of Biomechanical Engineering
dc.identifier.doi10.1115/1.4050642
dc.journal.fristpage081005-1
dc.journal.lastpage081005-9
dc.page9
dc.treeJournal of Biomechanical Engineering:;2021:;volume( 143 ):;issue: 008
dc.contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record