Uncertainty Estimation of Biomedical Oximeters: Proposal for an Experimental Test-Bench
View/ Open
Date
2021Author
Marchionni, Paolo
Nobile, Stefano
Carnielli, Virgilio P.
Scalise, Lorenzo
Metadata
Show full item recordAbstract
Paramagnetic oximeters are frequently used in clinical practice to measure and titrate the fraction of inspired oxygen (FiO2) administered to a hypoxic patient. Newborns are prone to several diseases leading to hypoxia but also to oxidative damage, therefore a precise measurement of administered oxygen is of paramount importance. In this study, we evaluated and compared the performance of four commonly used gas analyzers at a Neonatal Unit. A paramagnetic gas analyzer was considered the standard for reference. Gas mixing performed by blenders was also evaluated. We first assessed the stability and composition of O2 and medical air. A settling time study was conducted to estimate a steady-state. Results show how tested blenders underestimated the real measures when FiO2 was <50% and overestimated them when FiO2 ≥50%. In all the tested oximeters the FiO2 was underestimated by 1.18%, but the overall agreement with the gold standard was very good.