Comparison of Neutron-Gamma Separation Qualities of Various Organic Scintillation Materials and Liquid Scintillator LSB-200
View/ Open
Date
2020Author
Matěj, Zdeněk
Mravec, Filip
Jančár, Aleš
Košťál, Michal
Kučera, František
Cvachovec, František
Prenosil, Václav
Kopecký, Zdeněk
Čulen, Jiří
Pecák, Oldřich
Metadata
Show full item recordAbstract
In this work, we compare the pulse-shape discrimination (PSD) properties of newly developed liquid scintillator LSB-200, EJ-299-33A, BC-501A, stilbene, p-terphenyl, and Hidex Aqualight in neutron field generated by the LVR-15 reactor with silicon filter utilization. Pulses from the scintillators are processed by the Neutron-Gamma spectrometer. This spectrometric system with fast digitizer card contains two analog–digital converters with a resolution of 12 bits and sampling frequency 500 × 106 Hz. For photomultiplier linearity improvement, the active divider has been used. Measured data from scintillators have been processed using the integration method and compared. The results are presented.