    Verification of Manual Digitization Methods During Experimental Simulation of Knee Motion

    Thumbnail
    View/Open
    vvuq_005_03_031004.pdf (847.2Kb)
    Date
    2020
    Author
    Hargett, Zachary
    Gutierrez, Manuel
    Harman, Melinda
    Metadata
    Show full item record
    Abstract
    Cadaveric testing is a common approach for verifying mathematical algorithms in computational modeling. In models of total knee replacement (TKR) motion, model inputs commonly include rigid body motions defined using the Grood–Suntay spatial linkage between tibial and femoral components. This approach requires definition of coordinate systems for each rigid TKR component based on fiducial points, manual digitization of a point cloud within the experimental setup, and registration of the orientation relative to bone marker arrays. This study compared variability between two different manual point digitization methods (hand-held stylus and pivot tool each registered in an optical tracking system). This was accomplished by verifying the mathematical algorithm used to calculate the coordinate system from digitized points and quantifying the variability of the digitization methods in a simulated cadaver limb experimental setup. For the hand-held stylus method, the standard deviation was 0.50 mm for the origin and 1.31, 0.51, and 0.62 deg for the x–y–z axes, respectively. Required digitization of each rigid marker array created additional errors of 0.54 mm for the origin and 1.70, 1.66, and 0.98 deg for the x–y–z axes, respectively. For the pivot tool method, the standard deviation was 0.35 mm for the origin and 0.37, 1.27, and 1.24 deg for the x–y–z axes, respectively. In this experimental setup, the pivot tool was the better option for minimizing error while providing repeatable manual digitization of fiducial points and point clouds.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4275977
    Collections
    • Journal of Verification, Validation and Uncertainty Quantification

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV