    Improved Treatment of Postpartum Hemorrhage: Design, Development, and Bench-Top Validation of a Reusable Intrauterine Tamponade Device for Low-Resource Settings

    Thumbnail
    View/Open
    med_014_01_014503.pdf (984.7Kb)
    Date
    2020
    Author
    Hu, Katherine
    Lapinski, Maya M.
    Mischler, Gavin
    Allen, Robert H.
    Manbachi, Amir
    Seay, Rachel Chan
    Metadata
    Show full item record
    Abstract
    Postpartum hemorrhage (PPH) is the leading cause of maternal mortality worldwide, and effective interventions for addressing PPH are urgently needed. Uterine balloon tamponade (UBT) is a technique to control PPH. Commercially available UBT devices are expensive and frequently require imaging technology to ensure placement. Condom-catheter uterine tamponade (C-UBT) is a technique appropriate for low-resource settings. Testing of the C-UBT is needed to better understand and optimize this technique for use in a variety of clinical settings including low-resource contexts. We describe here the design, development, and bench-top validation of a reusable C-UBT device optimized for low-resource settings. The device was tested in three differently sized uterine models using a variety of condom balloon configurations. Intrauterine wall pressure application was measured to evaluate the device capacity to apply pressure of at least 90 mmHg, estimating the mean arterial pressure within the uterine vasculature. Bench-top experimental validation of pressure exerted in uterine models demonstrated the device's capability of reaching hemostatic pressure in uterine volumes ranging from 170 to 1740 mL. Device adaptability and versatility were shown through its ability to reach the target pressure of 90 mmHg in different uterine sizes by varying balloon parameters, including condom thickness and condom configuration. The results of this study show the potential of a low-cost, reusable C-UBT device optimized to treat PPH in a variety of clinical settings, including low-resource contexts.
    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4274340
    Collections
    • Journal of Medical Devices

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV
     

     

    Browse

    All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

    My Account

    LoginRegister

    Statistics

    View Usage Statistics

    DSpace software copyright © 2002-2015  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    @mire NV