Show simple item record

contributor authorA. F. Tencer
contributor authorM. C. Sherman
contributor authorK. D. Johnson
date accessioned2017-05-08T23:19:43Z
date available2017-05-08T23:19:43Z
date copyrightMay, 1985
date issued1985
identifier issn0148-0731
identifier otherJBENDY-25801#104_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/99531
description abstractIntramedullary rodding of femur fractures, although a safe and rapidly performed procedure, can result in several complications. If the rod fit is too loose, fracture instability, rod migration, and delayed union may result. If the rod fit is too tight, cracking of the femur may occur during rod insertion. These complications were investigated in terms of geometric and mechanical parameters of the bone-implant system. Results showed that rods of the same nominal size from different manufacturers showed more than twofold difference in flexural rigidity and a threefold difference in torsional modulus. These differences appear to be due to differences in cross sectional shape and wall thickness of the rods. Measurements of pushout force and hoop stress in cadaver femora showed a large difference in pushout force with different rods, and significantly lower forces in distal than in proximal femoral fracture components. Pushout force decreased with fracture component length proximally and dropped to zero in distal components less than 170 mm long. An increase in ream diameter in the distal components of just 1 mm was found to decrease the mean pushout force from 740N to 90N. The most significant variable was found to be anterior offset of the starting hole more than 6 mm from the centerline of the medullary canal which resulted in consistent lifting of the anterior cortex during insertion of the rod.
publisherThe American Society of Mechanical Engineers (ASME)
titleBiomechanical Factors Affecting Fracture Stability and Femoral Bursting in Closed Intramedullary Rod Fixation of Femur Fractures
typeJournal Paper
journal volume107
journal issue2
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.3138530
journal fristpage104
journal lastpage111
identifier eissn1528-8951
keywordsStability
keywordsFracture (Process)
keywordsBiomechanics
keywordsFemoral fractures
keywordsIntramedullary rods
keywordsForce
keywordsRods
keywordsShapes
keywordsStiffness
keywordsWall thickness
keywordsCanals
keywordsMeasurement
keywordsStress AND Bone
treeJournal of Biomechanical Engineering:;1985:;volume( 107 ):;issue: 002
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record