Show simple item record

contributor authorD. L. Grimes
contributor authorW. C. Flowers
contributor authorM. Donath
date accessioned2017-05-08T23:02:28Z
date available2017-05-08T23:02:28Z
date copyrightNovember, 1977
date issued1977
identifier issn0148-0731
identifier otherJBENDY-25540#215_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/89640
description abstractThe level walking process for an above knee (A/K ) amputee with a conventional prosthesis greatly compromises the amputee’s mobility. The fact that conventional prostheses lock in hyperextension during the stance phase, in contrast to the extend-flexextend pattern during stance for the natural limb, has been suggested as a source of the amputee’s uncosmetic gait and high energy expenditure. While the lock in hyperextension during stance provides stability to prevent buckling, it requires the person to vault over the prosthetic limb. This vaulting during level walking may cause higher vertical displacements of the body center of gravity (c.g.) and accompanying higher energy requirements for the amputee. This investigation employs an amputee-interactive prosthesis simulator system to evaluate the viability of controlling the prosthetic knee joint to follow a normal knee position pattern. In order to insure that the amputee-interactive prosthesis simulator system does not introduce gait anomalies, the system was controlled to simulate a conventional prosthesis. This showed that the simulator system has no undesired side effects since data from walking trials with the simulator system in “conventional prosthesis mode” are very similar to data from conventional prostheses in the literature. Then, an active position control scheme which controls the prosthetic knee joint to follow a normal knee position pattern was tested by two young, active amputees in level walking trials. The subjects experienced very little difficulty in walking with the active control scheme and preferred the simulator with the active control scheme to their conventional prostheses. Measured knee power requirements for the scheme indicate that this type of control is feasible without external power sources. However, measurements of the vertical displacement of the body c.g. show little difference between gait with the active control scheme and gait with a conventional prosthesis. It appears that the increased energy requirements for A/K amputees are not due in total to the lack of the extend-flex-extend position profile at the prosthetic knee joint.
publisherThe American Society of Mechanical Engineers (ASME)
titleFeasibility of an Active Control Scheme for Above Knee Prostheses
typeJournal Paper
journal volume99
journal issue4
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.3426293
journal fristpage215
journal lastpage221
identifier eissn1528-8951
keywordsStability
keywordsLocks (Waterways)
keywordsMeasurement
keywordsCenter of mass
keywordsProstheses
keywordsArtificial limbs
keywordsBuckling
keywordsDisplacement
keywordsPosition control
keywordsKnee AND Knee joint prostheses
treeJournal of Biomechanical Engineering:;1977:;volume( 099 ):;issue: 004
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)