description abstractA mathematical model is developed for calculating the pressures and flows in an artificial heart, its pneumatic drive unit, and a mock circulatory system. The system is divided into convenient subsystems to facilitate the analysis, and each subsystem is then analyzed separately. The set of independent equations developed is solved on a computer and corresponding experimental tests are made on the actual system. A comparison of the experimental and computer results shows good agreement for the mean flow rate through the pump and also for several instantaneous pressures and flow rates in the system.
publisherThe American Society of Mechanical Engineers (ASME)
titleAnalysis of an Artificial Ventricle and Mock Circulatory System
typeJournal Paper
journal volume99
journal issue4
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.3426288
journal fristpage184
journal lastpage188
identifier eissn1528-8951
keywordsCardiovascular system
keywordsArtificial ventricles
keywordsFlow (Dynamics)
keywordsComputers
keywordsPumps
keywordsEquations AND Artificial hearts
treeJournal of Biomechanical Engineering:;1977:;volume( 099 ):;issue: 004