Show simple item record

contributor authorWiebrecht, Jason J.
contributor authorStrick, Jacob A.
contributor authorFarris, Ryan J.
contributor authorSawicki, Jerzy T.
date accessioned2025-08-20T09:37:22Z
date available2025-08-20T09:37:22Z
date copyright4/28/2025 12:00:00 AM
date issued2025
identifier issn0148-0731
identifier otherbio_147_06_064501.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4308577
description abstractRobot-assisted gait rehabilitation is an increasingly common therapeutic intervention for enhancing locomotion and improving quality of life for children with lower-limb mobility impairments. However, there are few systems specifically designed for pediatric use, and those that do exist are largely cumbersome, bulky, and noncustom devices that ultimately reduce therapy effectiveness. This paper introduces the Cable-Driven Joint System (CDJS), a novel approach for pediatric gait rehabilitation that addresses these shortcomings in a lightweight and compact robotic device using the patient's professionally fitted orthosis. The CDJS consists of a 2.1 kg actuation unit that is held by a clinician which delivers assistive torques through a Bowden cable transmission to a 0.3 kg joint mounted to user-custom bracing. This work details an actuator benchtop evaluation, demonstrating a peak torque of 20 N·m, peak velocity of 7.2 rad/s, bandwidth of 9.7 Hz, and a mass moment of inertia of 58.38 kg cm2. An actuator model was developed and evaluated in simulation, showing a strong correlation with the experimental torque data (R-squared = 0.95) and indicating a transmission efficiency of 72%. In-air gait tracking experiments on an emulated subject showed that the CDJS assisted the subject to track a nominal knee trajectory with an average root-mean-squared error of 2.56 deg at a continuous torque of 1.37 N·m. These results suggest that the cable-driven actuator meets the design requirements for use in pediatric gait rehabilitation and is ready for implementation in clinical device trials.
publisherThe American Society of Mechanical Engineers (ASME)
titleDesign and Validation of a Cable-Driven Joint Actuator for Pediatric Knee Orthoses
typeJournal Paper
journal volume147
journal issue6
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4068369
journal fristpage64501-1
journal lastpage64501-9
page9
treeJournal of Biomechanical Engineering:;2025:;volume( 147 ):;issue: 006
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record