Show simple item record

contributor authorSantana, Gabriel
contributor authorHowell, Stephen M.
contributor authorHull, Maury L.
date accessioned2024-12-24T19:14:09Z
date available2024-12-24T19:14:09Z
date copyright3/20/2024 12:00:00 AM
date issued2024
identifier issn1932-6181
identifier othermed_018_01_011002.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4303554
description abstractDuring total knee arthroplasty (TKA) surgery, an important step is determining the correct insert thickness for each patient. If the insert is too thick, then stiffness results. Or conversely, if the insert is too thin, then instability results. One common method used to determine the insert thickness is by manually assessing the joint laxity; this is a qualitative method that depends on the surgeon's experience and “feel” and is unreliable. The lack of objective methods to reliably determine the correct insert thickness creates a need to develop such a method. One possible method is to measure the force required to push a trial insert into position, requiring a specialized tool to measure the push force. Hence, a new measuring tool was designed to measure the push force intra-operatively, accurately, and safely. To demonstrate functionality, the tool was tested on three patients. During the tests, the surgeon determined the appropriate thicknesses of the insert to trial and proceeded to position three different insert thicknesses ranging from 10 mm to 12 mm, and example push forces were recorded. The new tool met all the design criteria, and the example results from the patient testing show potential in using the peak push force to identify the insert with the correct thickness.
publisherThe American Society of Mechanical Engineers (ASME)
titleA New Handheld Tool to Select the Correct Tibial Insert Thickness by Measuring the Force to Push a Trial Insert Into Position During Total Knee Arthroplasty Surgery
typeJournal Paper
journal volume18
journal issue1
journal titleJournal of Medical Devices
identifier doi10.1115/1.4064973
journal fristpage11002-1
journal lastpage11002-5
page5
treeJournal of Medical Devices:;2024:;volume( 018 ):;issue: 001
contenttypeFulltext