Show simple item record

contributor authorBlauer, Joshua J. E.;Gray, Richard A.;Swenson, Darrell J.;Pathmanathan, Pras
date accessioned2023-04-06T13:03:09Z
date available2023-04-06T13:03:09Z
date copyright1/13/2023 12:00:00 AM
date issued2023
identifier issn23772158
identifier othervvuq_007_04_041007.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4288988
description abstractSurvival rates for sudden cardiac death treated with external defibrillation are estimated to be up to five times greater compared to cardiopulmonary resuscitation alone. Computational modeling can be used to investigate the relationship between patch location and defibrillation efficacy. However, credibility of model predictions is unclear. The aims of this paper are to (1) assess credibility of a commonly used computational approach for predicting impact of patch relocation on defibrillation efficacy; and (2) provide a concrete biomedical example of a model validation study with supporting applicability analysis, to systematically assess the relevance of the validation study for a proposed model context of use (COU). An electrostatic heart and torso computational model was developed. Simulations were compared against experimental recordings from a swine subject with external patches and multiple body surface and intracardiac recording electrodes. The applicability of this swine validation study to the human COU was assessed using an applicability analysis framework. Knowledge gaps identified by the applicability analysis were addressed using sensitivity analyses. In the swine validation study, quantitative agreement (R2 = 0.85) was observed between predicted and observed potentials at both surface and intracardiac electrodes using a leftright patch placement. Applicability analysis identified uncertainty in tissue conductivities as one of the main potential sources of unreliability; however, a sensitivity the analysis demonstrated that uncertainty in conductivity parameters had relatively little impact on model predictions (less than 10% relative change for twofold conductivity changes). We believe the results support pursuing human simulations further to evaluate impact of patch relocation.
publisherThe American Society of Mechanical Engineers (ASME)
titleValidation and Applicability Analysis of a Computational Model of External Defibrillation
typeJournal Paper
journal volume7
journal issue4
journal titleJournal of Verification, Validation and Uncertainty Quantification
identifier doi10.1115/1.4056596
journal fristpage41007
journal lastpage410078
page8
treeJournal of Verification, Validation and Uncertainty Quantification:;2023:;volume( 007 ):;issue: 004
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record