Show simple item record

contributor authorInacio, Jordan V.
contributor authorSchwarzenberg, Peter
contributor authorYoon, Richard S.
contributor authorKantzos, Andrew
contributor authorMalige, Ajith
contributor authorNwachuku, Chinenye O.
contributor authorDailey, Hannah L.
date accessioned2022-05-08T08:31:01Z
date available2022-05-08T08:31:01Z
date copyright3/11/2022 12:00:00 AM
date issued2022
identifier issn0148-0731
identifier otherbio_144_08_081009.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4284026
description abstractThe mechanics of distal femur fracture fixation has been widely studied in bench tests that employ a variety of approaches for holding and constraining femurs to apply loads. No standard test methods have been adopted for these tests and the impact of test setup on inferred construct mechanics has not been reported. Accordingly, the purpose of this study was to use finite element models to compare the mechanical performance of a supracondylar osteotomy with lateral plating under conditions that replicate several common bench test methods. A literature review was used to define a parameterized virtual model of a plated distal femur osteotomy in axial compression loading with four boundary condition sets ranging from minimally to highly constrained. Axial stiffness, fracture gap closure, and transverse motion at the fracture line were recorded for a range of applied loads and bridge spans. The results showed that construct mechanical performance was highly sensitive to boundary conditions imposed by the mechanical test fixtures. Increasing the degrees of constraint, for example, by potting and rigidly clamping one or more ends of the specimen, caused up to a 25× increase in axial stiffness of the construct. Transverse motion and gap closure at the fracture line, which is an important driver of interfragmentary strain, was also largely influenced by the constraint test setup. These results suggest that caution should be used when comparing reported results between bench tests that use different fixtures and that standardization of testing methods is needed in this field.
publisherThe American Society of Mechanical Engineers (ASME)
titleBoundary Conditions Matter—Impact of Test Setup on Inferred Construct Mechanics in Plated Distal Femur Osteotomies
typeJournal Paper
journal volume144
journal issue8
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4053875
journal fristpage81009-1
journal lastpage81009-11
page11
treeJournal of Biomechanical Engineering:;2022:;volume( 144 ):;issue: 008
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record