Show simple item record

contributor authorSwanson, Eric C.
contributor authorWeathersby, Ethan J.
contributor authorCagle, John C.
contributor authorSanders, Joan E.
date accessioned2019-09-18T09:00:52Z
date available2019-09-18T09:00:52Z
date copyright7/15/2019 12:00:00 AM
date issued2019
identifier issn0148-0731
identifier otherbio_141_10_101009
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4257885
description abstractUnderstanding the pressure distributions at the limb-socket interface is essential to the design and evaluation of prosthetic components for lower limb prosthesis users. Force sensing resistors (FSRs) are employed in prosthetics research to measure pressure at this interface due to their low cost, thin profile, and ease of use. While FSRs are known to be sensitive to many sources of error, few studies have systematically quantified these errors using test conditions relevant to lower limb prosthetics. The purpose of this study was to evaluate FSR accuracy for the measurement of lower limb prosthetics interface pressures. Two FSR models (Flexiforce A201 and Interlink 402) were subjected to a series of prosthetic-relevant tests. These tests included: (1) static compression, (2) cyclic compression, and (3) a combined static and cyclic compression protocol mimicking a variable activity (walk–sit–stand) procedure. Flexiforce sensors outperformed Interlink sensors and were then subjected to two additional tests: (4) static curvature and (5) static shear stress. Results demonstrated that FSRs experienced significant errors in all five tests. We concluded that: (1) if used carefully, FSRs can provide an estimate of prosthetic interface pressure, but these measurements should be interpreted within the expected range of possible measurement error given the setup; (2) FSRs should be calibrated in a setup that closely matches how they will be used for taking measurements; and (3) both Flexiforce and Interlink sensors can be used to estimate interface pressures; however, in most cases Flexiforce sensors are likely to provide more accurate measurements.
publisherAmerican Society of Mechanical Engineers (ASME)
titleEvaluation of Force Sensing Resistors for the Measurement of Interface Pressures in Lower Limb Prosthetics
typeJournal Paper
journal volume141
journal issue10
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4043561
journal fristpage101009
journal lastpage101009-13
treeJournal of Biomechanical Engineering:;2019:;volume( 141 ):;issue: 010
contenttypeFulltext