Show simple item record

contributor authorWu, Ying Ying
contributor authorAcharya, Deepshikha
contributor authorXu, Camilla
contributor authorCheng, Boyle
contributor authorRana, Sandeep
contributor authorShimada, Kenji
date accessioned2019-02-28T11:05:00Z
date available2019-02-28T11:05:00Z
date copyright7/13/2018 12:00:00 AM
date issued2018
identifier issn1932-6181
identifier othermed_012_03_031003.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252490
description abstractNoninvasive ventilator support using bi-level positive airway pressure/continuous positive airway pressure (BiPAP/CPAP) is commonly utilized for chronic medical conditions like sleep apnea and neuromuscular disorders like amyotrophic lateral sclerosis (ALS) that lead to weakness of respiratory muscles. Generic masks come in standard sizes and are often perceived by patients as being uncomfortable, ill-fitting, and leaky. A significant number of patients are unable to tolerate the masks and eventually stop using their devices. The goal of this project is to develop custom-fit masks to increase comfort, decrease air leakage, and thereby improve patient compliance. A single-patient case study of a patient with variant ALS was performed to evaluate the custom-fit masks. His high nose bridge and overbite of lower jaw caused poor fit with generic masks, and he was noncompliant with his machine. Using desktop Stereolithography three-dimensional (3D) printing and magnetic resonance imaging (MRI) data, a generic mask was extended with a rigid interface such that it was complementary to the patient's unique facial contours. Patient or clinicians interactively select a desired mask shape using a newly developed computer program. Subsequently, a compliant silicone layer was applied to the rigid interface. Ten different custom-fit mask designs were made using computer-aided design software. Patient evaluated the comfort, extent of leakage, and satisfaction of each mask via a questionnaire. All custom-fit masks were rated higher than the standard mask except for two. Our results suggest that modifying generic masks with a 3D-printed custom-fit interface is a promising strategy to improve compliance with BiPAP/CPAP machines.
publisherThe American Society of Mechanical Engineers (ASME)
titleCustom-Fit Three-Dimensional-Printed BiPAP Mask to Improve Compliance in Patients Requiring Long-Term Noninvasive Ventilatory Support
typeJournal Paper
journal volume12
journal issue3
journal titleJournal of Medical Devices
identifier doi10.1115/1.4040187
journal fristpage31003
journal lastpage031003-8
treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 003
contenttypeFulltext