    Performance of the Vaisala RS80A/H and RS90 Humicap Sensors and the Meteolabor “Snow White” Chilled-Mirror Hygrometer in Paramaribo, Suriname

    Source: Journal of Atmospheric and Oceanic Technology:;2006:;volume( 023 ):;issue: 011::page 1506
    Author:
    Verver, Gé
    ,
    Fujiwara, Masatomo
    ,
    Dolmans, Pier
    ,
    Becker, Cor
    ,
    Fortuin, Paul
    ,
    Miloshevich, Larry
    DOI: 10.1175/JTECH1941.1
    Publisher: American Meteorological Society
    Abstract: In climate research there is a strong need for accurate observations of water vapor in the upper atmosphere. Radiosoundings provide relative humidity profiles but the accuracy of many routine instruments is notoriously inadequate in the cold upper troposphere. In this study results from a soundings program executed in Paramaribo, Suriname (5.8°N, 55.2°W), are presented. The aim of this program was to compare the performance of different humidity sensors in the upper troposphere in the Tropics and to test different bias corrections suggested in the literature. The payload of each sounding consisted of a chilled-mirror ?Snow White? sensor from Meteolabor AG, which was used as a reference, and two additional sensors from Vaisala, that is, either the RS80A, the RS80H, or the RS90. In total 37 separate soundings were made. For the RS80A a clear, dry bias of between ?4% and ?8% RH is found in the lower troposphere compared to the Snow White observation, confirming the findings in previous studies. A mean dry bias was found in the upper troposphere, which could be effectively corrected. The RS80H sensor shows a significant wet bias of 2%?5% in RH in the middle and upper troposphere, which has not been reported before. Comparing observations with RS80H sensors of different ages gives no indication of sensor aging or sensor contamination. It is therefore concluded that the plastic cover introduced by Vaisala to avoid sensor contamination is effective. Finally, the RS90 sensor yields a small but significant wet bias of 2%?3% below 7-km altitude. The time-lag error correction from Miloshevich et al. was applied to the Vaisala data, which resulted in an increased variability in the relative humidity profile above 9- (RS80A), 8- (RS80H), and 11-km (RS90) altitude, respectively, which is in better agreement with the Snow White data. The averaged Snow White profile is compared with the average profiles of relative humidity from the European Centre for Medium-Range Weather Forecasts (ECMWF). No significant bias is found in either the analyses or the forecasts. The correlation coefficient for the Snow White and ECMWF data between 200 and 800 hPa was 0.66 for the 36-h forecast and 0.77 for the analysis.
    • Download: (1.522Mb)
    • Show Full MetaData Hide Full MetaData
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Performance of the Vaisala RS80A/H and RS90 Humicap Sensors and the Meteolabor “Snow White” Chilled-Mirror Hygrometer in Paramaribo, Suriname

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4227649
    Collections
    • Journal of Atmospheric and Oceanic Technology

    Show full item record

    contributor authorVerver, Gé
    contributor authorFujiwara, Masatomo
    contributor authorDolmans, Pier
    contributor authorBecker, Cor
    contributor authorFortuin, Paul
    contributor authorMiloshevich, Larry
    date accessioned2017-06-09T17:23:20Z
    date available2017-06-09T17:23:20Z
    date copyright2006/11/01
    date issued2006
    identifier issn0739-0572
    identifier otherams-84325.pdf
    identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4227649
    description abstractIn climate research there is a strong need for accurate observations of water vapor in the upper atmosphere. Radiosoundings provide relative humidity profiles but the accuracy of many routine instruments is notoriously inadequate in the cold upper troposphere. In this study results from a soundings program executed in Paramaribo, Suriname (5.8°N, 55.2°W), are presented. The aim of this program was to compare the performance of different humidity sensors in the upper troposphere in the Tropics and to test different bias corrections suggested in the literature. The payload of each sounding consisted of a chilled-mirror ?Snow White? sensor from Meteolabor AG, which was used as a reference, and two additional sensors from Vaisala, that is, either the RS80A, the RS80H, or the RS90. In total 37 separate soundings were made. For the RS80A a clear, dry bias of between ?4% and ?8% RH is found in the lower troposphere compared to the Snow White observation, confirming the findings in previous studies. A mean dry bias was found in the upper troposphere, which could be effectively corrected. The RS80H sensor shows a significant wet bias of 2%?5% in RH in the middle and upper troposphere, which has not been reported before. Comparing observations with RS80H sensors of different ages gives no indication of sensor aging or sensor contamination. It is therefore concluded that the plastic cover introduced by Vaisala to avoid sensor contamination is effective. Finally, the RS90 sensor yields a small but significant wet bias of 2%?3% below 7-km altitude. The time-lag error correction from Miloshevich et al. was applied to the Vaisala data, which resulted in an increased variability in the relative humidity profile above 9- (RS80A), 8- (RS80H), and 11-km (RS90) altitude, respectively, which is in better agreement with the Snow White data. The averaged Snow White profile is compared with the average profiles of relative humidity from the European Centre for Medium-Range Weather Forecasts (ECMWF). No significant bias is found in either the analyses or the forecasts. The correlation coefficient for the Snow White and ECMWF data between 200 and 800 hPa was 0.66 for the 36-h forecast and 0.77 for the analysis.
    publisherAmerican Meteorological Society
    titlePerformance of the Vaisala RS80A/H and RS90 Humicap Sensors and the Meteolabor “Snow White” Chilled-Mirror Hygrometer in Paramaribo, Suriname
    typeJournal Paper
    journal volume23
    journal issue11
    journal titleJournal of Atmospheric and Oceanic Technology
    identifier doi10.1175/JTECH1941.1
    journal fristpage1506
    journal lastpage1518
    treeJournal of Atmospheric and Oceanic Technology:;2006:;volume( 023 ):;issue: 011
    contenttypeFulltext