    Archive

    Realistic Uncertainties in Pressure and Temperature Calibration Reference Values

    Source: Journal of Atmospheric and Oceanic Technology:;1984:;volume( 001 ):;issue: 002::page 115
    Author:
    Pike, Julian M.
    DOI: 10.1175/1520-0426(1984)001<0115:RUIPAT>2.0.CO;2
    Publisher: American Meteorological Society
    Abstract: Field instruments require calibration against standards that introduce-insignificant errors of their own in the process. This paper discusses a pressure and temperature calibration facility, analyzes the error levels attained in its routine use and shows that the accuracies are both reasonable and adequate. The results are given in terms of uncertainty; values of 0.025 kPa for pressure and 0.040 K for temperature were shown to meet the needs for calibration of research quality sensors.
    • Download: (390.8Kb)
    • Show Full MetaData Hide Full MetaData
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      Realistic Uncertainties in Pressure and Temperature Calibration Reference Values

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4211733
    Collections
    • Journal of Atmospheric and Oceanic Technology

    Show full item record

    contributor authorPike, Julian M.
    date accessioned2017-06-09T16:33:39Z
    date available2017-06-09T16:33:39Z
    date copyright1984/06/01
    date issued1984
    identifier issn0739-0572
    identifier otherams-70.pdf
    identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4211733
    description abstractField instruments require calibration against standards that introduce-insignificant errors of their own in the process. This paper discusses a pressure and temperature calibration facility, analyzes the error levels attained in its routine use and shows that the accuracies are both reasonable and adequate. The results are given in terms of uncertainty; values of 0.025 kPa for pressure and 0.040 K for temperature were shown to meet the needs for calibration of research quality sensors.
    publisherAmerican Meteorological Society
    titleRealistic Uncertainties in Pressure and Temperature Calibration Reference Values
    typeJournal Paper
    journal volume1
    journal issue2
    journal titleJournal of Atmospheric and Oceanic Technology
    identifier doi10.1175/1520-0426(1984)001<0115:RUIPAT>2.0.CO;2
    journal fristpage115
    journal lastpage119
    treeJournal of Atmospheric and Oceanic Technology:;1984:;volume( 001 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian