    The Calibration and Intercalibration of Sea-Going Infrared Radiometer Systems Using a Low Cost Blackbody Cavity

    Source: Journal of Atmospheric and Oceanic Technology:;1999:;volume( 016 ):;issue: 009::page 1183
    Author:
    Donlon, C. J.
    ,
    Nightingale, T.
    ,
    Fiedler, L.
    ,
    Fisher, G.
    ,
    Baldwin, D.
    ,
    Robinson, I. S.
    DOI: 10.1175/1520-0426(1999)016<1183:TCAIOS>2.0.CO;2
    Publisher: American Meteorological Society
    Abstract: There are many infrared radiometer systems available for the measurement of in situ sea surface skin temperature (SSST). Unfortunately, the marine environment is extremely hostile to optical components, and to ensure the accuracy of SSST measurements, an absolute calibration of instrumentation using an independent calibration reference is required both before and after any sea deployment. During extended deployments it is prudent to have additional regular calibration data to monitor instrument performance characteristics. This paper presents a design for an ambient temperature (278?325 K), wide aperture (100 mm), reference blackbody unit that may be used to calibrate a variety of sea-going infrared radiometer systems both in the laboratory and in the field. The blackbody consists of a spun copper cavity coated with well-characterized high emissivity paint (Mankiewicz Nextel Velvet Coating 811-21) immersed in a water bath that is continuously mixed using a strong water pump. The radiant temperature of the blackbody cavity is determined from the measured water bath temperature. Results derived from validation and intercomparison experiments show this blackbody design to be an accurate and reliable reference blackbody source. However, in order to ensure that the best possible calibration data are obtained, extreme care must be taken to ensure the accurate measurement of the water bath temperature, proper positioning of a radiometer in front of the cavity itself, and prevention of condensation on the cavity surface. Four blackbody units have been specifically built for the European Union combined action for the study of the ocean thermal skin (CASOTS) program. Using these units as reference radiance sources, the authors describe the strategy adopted and present results obtained from the CASOTS radiometer intercalibration experiment. These results highlight the need to obtain independent calibration data both before and after sea-going radiometer deployments and the need to standardize field radiometer calibration protocols.
    • Download: (205.8Kb)
    • Show Full MetaData Hide Full MetaData
    • Item Order
    • Go To Publisher
    • Price: 5000 Rial
    • Statistics

      The Calibration and Intercalibration of Sea-Going Infrared Radiometer Systems Using a Low Cost Blackbody Cavity

    URI
    http://yetl.yabesh.ir/yetl1/handle/yetl/4151523
    Collections
    • Journal of Atmospheric and Oceanic Technology

    Show full item record

    contributor authorDonlon, C. J.
    contributor authorNightingale, T.
    contributor authorFiedler, L.
    contributor authorFisher, G.
    contributor authorBaldwin, D.
    contributor authorRobinson, I. S.
    date accessioned2017-06-09T14:15:25Z
    date available2017-06-09T14:15:25Z
    date copyright1999/09/01
    date issued1999
    identifier issn0739-0572
    identifier otherams-1581.pdf
    identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4151523
    description abstractThere are many infrared radiometer systems available for the measurement of in situ sea surface skin temperature (SSST). Unfortunately, the marine environment is extremely hostile to optical components, and to ensure the accuracy of SSST measurements, an absolute calibration of instrumentation using an independent calibration reference is required both before and after any sea deployment. During extended deployments it is prudent to have additional regular calibration data to monitor instrument performance characteristics. This paper presents a design for an ambient temperature (278?325 K), wide aperture (100 mm), reference blackbody unit that may be used to calibrate a variety of sea-going infrared radiometer systems both in the laboratory and in the field. The blackbody consists of a spun copper cavity coated with well-characterized high emissivity paint (Mankiewicz Nextel Velvet Coating 811-21) immersed in a water bath that is continuously mixed using a strong water pump. The radiant temperature of the blackbody cavity is determined from the measured water bath temperature. Results derived from validation and intercomparison experiments show this blackbody design to be an accurate and reliable reference blackbody source. However, in order to ensure that the best possible calibration data are obtained, extreme care must be taken to ensure the accurate measurement of the water bath temperature, proper positioning of a radiometer in front of the cavity itself, and prevention of condensation on the cavity surface. Four blackbody units have been specifically built for the European Union combined action for the study of the ocean thermal skin (CASOTS) program. Using these units as reference radiance sources, the authors describe the strategy adopted and present results obtained from the CASOTS radiometer intercalibration experiment. These results highlight the need to obtain independent calibration data both before and after sea-going radiometer deployments and the need to standardize field radiometer calibration protocols.
    publisherAmerican Meteorological Society
    titleThe Calibration and Intercalibration of Sea-Going Infrared Radiometer Systems Using a Low Cost Blackbody Cavity
    typeJournal Paper
    journal volume16
    journal issue9
    journal titleJournal of Atmospheric and Oceanic Technology
    identifier doi10.1175/1520-0426(1999)016<1183:TCAIOS>2.0.CO;2
    journal fristpage1183
    journal lastpage1197
    treeJournal of Atmospheric and Oceanic Technology:;1999:;volume( 016 ):;issue: 009
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian