Show simple item record

contributor authorHariharan, Prasanna
contributor authorD’Souza, Gavin
contributor authorHorner, Marc
contributor authorMalinauskas, Richard A.
contributor authorMyers, Matthew R.
date accessioned2017-05-09T01:15:24Z
date available2017-05-09T01:15:24Z
date issued2015
identifier issn0148-0731
identifier otherbio_137_09_094501.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/157182
description abstractAs part of an ongoing effort to develop verification and validation (V&V) standards for using computational fluid dynamics (CFD) in the evaluation of medical devices, we have developed idealized flowbased verification benchmarks to assess the implementation of commonly cited powerlaw based hemolysis models in CFD. The verification process ensures that all governing equations are solved correctly and the model is free of user and numerical errors. To perform verification for powerlaw based hemolysis modeling, analytical solutions for the Eulerian powerlaw blood damage model (which estimates hemolysis index (HI) as a function of shear stress and exposure time) were obtained for Couette and inclined Couette flow models, and for Newtonian and nonNewtonian pipe flow models. Subsequently, CFD simulations of fluid flow and HI were performed using Eulerian and three different Lagrangianbased hemolysis models and compared with the analytical solutions. For all the geometries, the blood damage results from the Eulerianbased CFD simulations matched the Eulerian analytical solutions within ∼1%, which indicates successful implementation of the Eulerian hemolysis model. Agreement between the Lagrangian and Eulerian models depended upon the choice of the hemolysis powerlaw constants. For the commonly used values of powerlaw constants (خ±  = 1.9–2.42 and خ²  = 0.65–0.80), in the absence of flow acceleration, most of the Lagrangian models matched the Eulerian results within 5%. In the presence of flow acceleration (inclined Couette flow), moderate differences (∼10%) were observed between the Lagrangian and Eulerian models. This difference increased to greater than 100% as the beta exponent decreased. These simplified flow problems can be used as standard benchmarks for verifying the implementation of blood damage predictive models in commercial and opensource CFD codes. The current study used only a powerlaw model as an illustrative example to emphasize the need for model verification. Similar verification problems could be developed for other types of hemolysis models (such as strainbased and energy dissipationbased methods). And since the current study did not include experimental validation, the results from the verified models do not guarantee accurate hemolysis predictions. This verification step must be followed by experimental validation before the hemolysis models can be used for actual device safety evaluations.
publisherThe American Society of Mechanical Engineers (ASME)
titleVerification Benchmarks to Assess the Implementation of Computational Fluid Dynamics Based Hemolysis Prediction Models
typeJournal Paper
journal volume137
journal issue9
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4030823
journal fristpage94501
journal lastpage94501
identifier eissn1528-8951
treeJournal of Biomechanical Engineering:;2015:;volume( 137 ):;issue: 009
contenttypeFulltext


Files in this item