Show simple item record

contributor authorJingjing, Xiao
contributor authorLong, Huang
contributor authorLijun, Shen
contributor authorYang, Yang
date accessioned2017-05-09T01:11:23Z
date available2017-05-09T01:11:23Z
date issued2014
identifier issn1932-6181
identifier othermed_008_04_044501.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/155970
description abstractWith reference to the study of the robotic system for ophthalmology microsurgery, a robotic system was presented with the key mechanism design and control subsystem scheme considering the procedures and features of retinal vascular bypass surgery. After discussion of the clinical application, the humanmachine cooperated surgical process of the robotic system for retinal vascular bypass surgery was described, and the design of the surgical layout environment was proposed. The intraocular motion planning method was also described and corresponding experiments using a table tennis ball model and an in vitro porcine eye model were performed. Finally, the feasibility analysis and robotic system error analysis were also summarized and future works discussed for further research.
publisherThe American Society of Mechanical Engineers (ASME)
titleDesign and Research of a Robotic Aided System for Retinal Vascular Bypass Surgery
typeJournal Paper
journal volume8
journal issue4
journal titleJournal of Medical Devices
identifier doi10.1115/1.4027230
journal fristpage44501
journal lastpage44501
identifier eissn1932-619X
treeJournal of Medical Devices:;2014:;volume( 008 ):;issue: 004
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record