Show simple item record

contributor authorEmily Paukert
contributor authorSusan Mantell
contributor authorJohn Korkko
contributor authorBruce Hammer
contributor authorPhil Williams
date accessioned2017-05-09T00:46:13Z
date available2017-05-09T00:46:13Z
date copyrightJune, 2011
date issued2011
identifier issn1932-6181
identifier otherJMDOA4-28018#027538_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/147258
description abstractAll materials experience a force when placed in a region of magnetic field and field gradient. The magnitude of this force depends on the magnetic susceptibility of the material and this varies over a wide range depending on the type of material. Our goal is to develop a technique for evaluating the magnetic susceptibility of cells and subcellular organelles so that scientists can access to develop new methods to modify or modulate internal cellular forces. Research studies have shown that forces in the piconewton range can affect cellular behavior. Internal forces of this magnitude can occur in cells exposed to high intensity magnetic fields, if the difference in magnetic susceptibility of subcellular organelles is as low as 10%. Because the magnetic susceptibility x is expected to be on the order of 9×10−6, the proposed measurement technique must be extremely sensitive. In this paper, a pilot study is described in which the feasibility of a magnetophoresis technique is explored. Tests implementing magnetophoresis for polystyrene test particles (|x|=8.21×10−6) with a 100 μm diameter explored the sensitivity and accuracy effects of varying fluid flow speeds of 0.63 mm/s, 1.09 mm/s, and 1.44 mm/s, particle radius to channel depth ratios (r/a) of 0.043 and 0.199, and a magnetic field and gradient product (B∗dB/dz) of 38.91 T2/m. The percent uncertainties of the experimental magnetic susceptibilities for the three different flow speeds and r/a ratio combinations studied are 12.3%, 18.3%, and 22.4% (in order of flow speed). The trial runs indicate that a balance of a larger r/a ratio and a slower flow speed is ideal to optimize consistency in flow velocities and calculated magnetic susceptibilities while minimizing uncertainty. Requirements for MEMs device design are also presented.
publisherThe American Society of Mechanical Engineers (ASME)
titleFeasibility Study of MEMs Technique for Characterizing Magnetic Susceptibility of Subcellular Organelles
typeJournal Paper
journal volume5
journal issue2
journal titleJournal of Medical Devices
identifier doi10.1115/1.3591391
journal fristpage27538
identifier eissn1932-619X
treeJournal of Medical Devices:;2011:;volume( 005 ):;issue: 002
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record