Show simple item record

contributor authorI. Zivkovic
contributor authorM. Gonzalez
contributor authorF. Amirouche
date accessioned2017-05-09T00:36:40Z
date available2017-05-09T00:36:40Z
date copyrightApril, 2010
date issued2010
identifier issn0148-0731
identifier otherJBENDY-27127#041008_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/142638
description abstractThis paper explores the effect of under-reaming on micromotion at the cup/bone interface of a press-fit acetabular cup. A cadaver experiment was performed on 11 acetabuli implanted with a cementless acetabular cup. The loading profile simulated hip impingement at the extremes of motion and subluxation relocation of the hip joint. Micromotion of each cup was measured in a custom made jig with linear variable differential transducers. A CAT scan and DEXA scan of the acetabulum and femoral head respectively were used to construct a three-dimensional patient specific finite element model of the hemi-pelvis. The model predicted cup micromotion under loading conditions and stresses in the acetabulum as a result of cup insertion. Micromotion was then calculated as a function of variable bone density and variable degree of underreaming. Simulated cup insertion with under-reaming of 2 mm or more approached or exceeded the yield strength of bone in acetabula with reduced bone mass density.
publisherThe American Society of Mechanical Engineers (ASME)
titleThe Effect of Under-Reaming on the Cup/Bone Interface of a Press Fit Hip Replacement
typeJournal Paper
journal volume132
journal issue4
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.2913228
journal fristpage41008
identifier eissn1528-8951
keywordsStress
keywordsBone
keywordsPresses
keywordsDensity AND Finite element model
treeJournal of Biomechanical Engineering:;2010:;volume( 132 ):;issue: 004
contenttypeFulltext