Show simple item record

contributor authorJoshua A. Nordquist
contributor authorM. L. Hull
date accessioned2017-05-09T00:31:48Z
date available2017-05-09T00:31:48Z
date copyrightMarch, 2009
date issued2009
identifier issn0148-0731
identifier otherJBENDY-26901#034505_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/140008
description abstractBecause instrumented spatial linkages (ISLs) have been commonly used in measuring joint rotations and must be calibrated before using the device in confidence, a calibration device design and associated method for quantifying calibration device error would be useful. The objectives of the work reported by this paper were to (1) design an ISL calibration device and demonstrate the design for a specific application, (2) describe a new method for calibrating the device that minimizes measurement error, and (3) quantify measurement error of the device using the new method. Relative translations and orientations of the device were calculated via a series of transformation matrices containing inherent fixed and variable parameters. These translations and orientations were verified with a coordinate measurement machine, which served as a gold standard. Inherent fixed parameters of the device were optimized to minimize measurement error. After parameter optimization, accuracy was determined. The root mean squared error (RMSE) was 0.175 deg for orientation and 0.587 mm for position. All RMSE values were less than 0.8% of their respective full-scale ranges. These errors are comparable to published measurement errors of ISLs for positions and lower by at least a factor of 2 for orientations. These errors are in spite of the many steps taken in design and manufacturing to achieve high accuracy. Because it is challenging to achieve the accuracy required for a custom calibration device to serve as a viable gold standard, it is important to verify that a calibration device provides sufficient precision to calibrate an ISL.
publisherThe American Society of Mechanical Engineers (ASME)
titleDesign and Evaluation of a New General-Purpose Device for Calibrating Instrumented Spatial Linkages
typeJournal Paper
journal volume131
journal issue3
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.2965375
journal fristpage34505
identifier eissn1528-8951
keywordsLinkages
keywordsDesign
keywordsCalibration
keywordsErrors AND Optimization
treeJournal of Biomechanical Engineering:;2009:;volume( 131 ):;issue: 003
contenttypeFulltext