Show simple item record

contributor authorW. Y. Gu
contributor authorM. A. Justiz
date accessioned2017-05-09T00:06:42Z
date available2017-05-09T00:06:42Z
date copyrightDecember, 2002
date issued2002
identifier issn0148-0731
identifier otherJBENDY-26278#790_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/126329
description abstractThis paper describes a new apparatus and method for measuring swelling dependent electrical conductivity of charged hydrated soft tissues. The apparatus was calibrated using a conductivity standard. Swelling dependent specific conductivity of porcine annulus fibrosis (AF) samples was determined. The conductivity values for porcine AF were similar to those for human and bovine articular cartilage found in the literature. Results revealed a significant linear correlation between specific conductivity and water content for porcine AF tissues tested in phosphate buffered saline (PBS).
publisherThe American Society of Mechanical Engineers (ASME)
titleApparatus for Measuring the Swelling Dependent Electrical Conductivity of Charged Hydrated Soft Tissues
typeJournal Paper
journal volume124
journal issue6
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.1516571
journal fristpage790
journal lastpage793
identifier eissn1528-8951
keywordsElectrical conductivity
keywordsBiological tissues
keywordsConductivity
keywordsWater
keywordsSoft tissues AND Cartilage
treeJournal of Biomechanical Engineering:;2002:;volume( 124 ):;issue: 006
contenttypeFulltext