description abstractAn electromagnetic flowmeter was developed to measure transient flows at a data rate of 60 Hz. The approach taken was to develop suitable electronics to replace the electronics of a commercially available electromagnetic flowmeter normally used for steady-state operation. Use of the commercially available flowmeter body, which includes the magnetic coils, core, and signal electrodes, provided a relatively economical means of fabricating the transient flowmeter. A transient flow calibration facility consisting of a free-falling water column was also designed and built. Results of the calibrations are presented and show that the flowmeter can accurately measure transient flows up to the maximum observed acceleration of approximately 1 g.
publisherThe American Society of Mechanical Engineers (ASME)
titleA Transient Electromagnetic Flowmeter and Calibration Facility
typeJournal Paper
journal volume112
journal issue1
journal titleJournal of Fluids Engineering
identifier doi10.1115/1.2909360
journal fristpage12
journal lastpage15
identifier eissn1528-901X
keywordsCalibration
keywordsFlowmeters
keywordsFlow (Dynamics)
keywordsElectronics
keywordsElectrodes
keywordsSignals
keywordsSteady state AND Water
treeJournal of Fluids Engineering:;1990:;volume( 112 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record