Show simple item record

contributor authorMichel Y. Jaffrin
contributor authorLuhui Ding
contributor authorJean Marc Laurent
date accessioned2017-05-08T23:32:07Z
date available2017-05-08T23:32:07Z
date copyrightMay, 1990
date issued1990
identifier issn0148-0731
identifier otherJBENDY-25858#212_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/106597
description abstractThe mass transfer in a hemodialyser in the presence of combined dialysis and ultrafiltration has been calculated by integration of mass fluxes across the boundary layers in blood and dialysate phase taking into account the partial rejection of solute as well as changes in local blood flow and ultrafiltration flux along the membrane. Clearances of creatinin, vitamin B12, and myoglobin have been calculated as a function of blood and ultrafiltrate flow rate and were found to be in good agreement with in vitro measurements. The data suggest the following empirical correlation for the hemodiafiltration clearance CL = CLD + 0.46QF where Q F is the ultrafiltration flow rate and C LD is the clearance at Q F =0.
publisherThe American Society of Mechanical Engineers (ASME)
titleSimultaneous Convective and Diffusive Mass Transfers in a Hemodialyser
typeJournal Paper
journal volume112
journal issue2
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.2891174
journal fristpage212
journal lastpage219
identifier eissn1528-8951
keywordsMass transfer
keywordsFlow (Dynamics)
keywordsClearances (Engineering)
keywordsBlood
keywordsBoundary layers
keywordsMembranes
keywordsBlood flow
keywordsMeasurement AND Flux (Metallurgy)
treeJournal of Biomechanical Engineering:;1990:;volume( 112 ):;issue: 002