Show simple item record

contributor authorSavio L.-Y. Woo
contributor authorThay Q. Lee
contributor authorPeter O. Newton
contributor authorMichael I. Danto
contributor authorKaren J. Ohland
date accessioned2017-05-08T23:32:01Z
date available2017-05-08T23:32:01Z
date copyrightNovember, 1990
date issued1990
identifier issn0148-0731
identifier otherJBENDY-25864#426_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/106547
description abstractDetermination of the tensile stresses in ligaments and tendons during uniaxial loading depends on accurate measurement of the cross-sectional area. In this study, a laser micrometer system was employed to evaluate the cross-sectional shape and area of the medial collateral ligament (MCL) at three locations and anterior cruciate ligament (ACL). In a New Zealand White (NZW) rabbit, morphologic sections of the ligaments were made to verify the cross-sectional shape reconstructed by the laser micrometer system. The areas obtained by the laser micrometer system from ten additional NZW rabbits were compared with those obtained by two other methods commonly used to measure the cross-sectional area of ligaments and tendons: one method uses digital calipers and the other a constant pressure (0.12 MPa) area micrometer. For the MCL, the digital calipers yielded results very similar to those of the laser micrometer, but the constant pressure area micrometer yielded values 20 percent lower. The area measured at the proximal site of the MCL was 13 percent greater than the area measured at the joint line and distal line. For the ACL, the values obtained by the digital calipers and constant pressure area micrometer were 16 and 20 percent lower, respectively. Because of the irregular shape exhibited by the rabbit ACL, the digital calipers could not accurately measure the crosssectional area. The constant pressure area micrometer yielded lower values for the cross-sectional area of both the MCL and ACL, presumably due to the applied pressure which caused changes in both the cross-sectional shape and area.
publisherThe American Society of Mechanical Engineers (ASME)
titleThe Use of a Laser Micrometer System to Determine the Cross-Sectional Shape and Area of Ligaments: A Comparative Study With Two Existing Methods
typeJournal Paper
journal volume112
journal issue4
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.2891206
journal fristpage426
journal lastpage431
identifier eissn1528-8951
keywordsLasers
keywordsShapes
keywordsPressure
keywordsAnterior cruciate ligament
keywordsTendons AND Tension
treeJournal of Biomechanical Engineering:;1990:;volume( 112 ):;issue: 004
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record