    Design of a Novel Insertion Training System for Central Venous Catheterization

    Source: Journal of Medical Devices:;2026:;volume( 020 ):;issue:005::page 1123
    Author:
    Worrall, William
    ,
    Moore, Jason Z.
    DOI: 10.1115/1.4071877
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Abstract. Central venous catheterization (CVC) is a common medical procedure with a complication rate of over 15%. Current CVC manikin-based simulators lack anatomical variation and objective feedback of user interaction skills, which greatly hinders effective skill acquisition. To address these limitations, a novel CVC insertion training system was developed and integrated with the Advanced Dynamic Haptic Robotic Trainer (DHRT+). This insertion training system contains a sensorized syringe, custom tissue phantom, and a tool detection subsystem, allowing residents to practice nearly the entire CVC procedure with realistic haptics and automated feedback on subcutaneous tool positioning. Five experiments were conducted to assess the haptic fidelity of the sensorized syringe and custom tissue and to evaluate the accuracy of subcutaneous tool detection. Results showed that the sensorized syringe, which detects aspiration and simulates arterial and venous flash, accurately reproduced clinical aspiration forces. The custom tissue phantom provided realistic force feedback during ultrasound (US) compression and needle insertion, comparable to that of commercial tissue phantoms at a substantially lower cost. The color sensors used to distinguish between inserted guidewires and catheters accurately verified tool depth and enabled tool insertion across multiple positions on the tissue surface. Overall, novel CVC insertion training system components were presented that enabled CVC simulation and provided accurate measurement on aspiration, guidewire depth, and catheter depth.
    • Download: (1.566Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Design of a Novel Insertion Training System for Central Venous Catheterization

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4315602
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorWorrall, William
    contributor authorMoore, Jason Z.
    date accessioned2026-08-23T07:47:11Z
    date available2026-08-23T07:47:11Z
    date copyright2026/10/01
    date issued2026
    identifier issn1932-6181
    identifier othermed-25-1253.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4315602
    description abstractAbstract. Central venous catheterization (CVC) is a common medical procedure with a complication rate of over 15%. Current CVC manikin-based simulators lack anatomical variation and objective feedback of user interaction skills, which greatly hinders effective skill acquisition. To address these limitations, a novel CVC insertion training system was developed and integrated with the Advanced Dynamic Haptic Robotic Trainer (DHRT+). This insertion training system contains a sensorized syringe, custom tissue phantom, and a tool detection subsystem, allowing residents to practice nearly the entire CVC procedure with realistic haptics and automated feedback on subcutaneous tool positioning. Five experiments were conducted to assess the haptic fidelity of the sensorized syringe and custom tissue and to evaluate the accuracy of subcutaneous tool detection. Results showed that the sensorized syringe, which detects aspiration and simulates arterial and venous flash, accurately reproduced clinical aspiration forces. The custom tissue phantom provided realistic force feedback during ultrasound (US) compression and needle insertion, comparable to that of commercial tissue phantoms at a substantially lower cost. The color sensors used to distinguish between inserted guidewires and catheters accurately verified tool depth and enabled tool insertion across multiple positions on the tissue surface. Overall, novel CVC insertion training system components were presented that enabled CVC simulation and provided accurate measurement on aspiration, guidewire depth, and catheter depth.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign of a Novel Insertion Training System for Central Venous Catheterization
    typeJournal Paper
    journal volume20
    journal issue5
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4071877
    journal fristpage1123
    journal lastpage1133
    page11
    treeJournal of Medical Devices:;2026:;volume( 020 ):;issue:005
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian