    Structure Design and Experimental Study on a Shape Memory Alloy Based Negative Pressure Sampling Capsule

    Source: Journal of Medical Devices:;2026:;volume( 020 ):;issue:002
    Author:
    Yang, Tianye
    ,
    Wang, Jiaole
    ,
    Song, Shuang
    DOI: 10.1115/1.4070417
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Abstract. This study aims to introduce a wireless, remotely triggered biopsy capsule based on shape memory alloy (SMA). This capsule utilizes the negative pressure suction principle to achieve liquid sampling within the digestive tract. The proposed capsule is designed with an SMA spring-driven sampling trigger mechanism. An external magnetic field directs the capsule to the sampling area and aligns the sampling port with the sampling liquid. A Bluetooth signal is then sent and energizes the SMA spring, causing deformation and generating negative pressure for liquid sampling. The capsule prototype has dimensions of ø 15.0 mm × 30.0 mm. Both simulation and in vitro experiments have been carried out. Results showed that the proposed capsule has a rapid response time of 1.4 s and can sample liquids with viscosities ranging from 0.7 mPa·s to 100 mPa·s with the maximum sampling volume around 0.2 mL. Furthermore, the capsule's sealing performance has also been verified. The experimental results indicate that the proposed capsule can achieve Bluetooth remote-triggered liquid sampling in the digestive tract and satisfy the clinical requirements.
    • Download: (1.659Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Structure Design and Experimental Study on a Shape Memory Alloy Based Negative Pressure Sampling Capsule

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4315570
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorYang, Tianye
    contributor authorWang, Jiaole
    contributor authorSong, Shuang
    date accessioned2026-08-23T07:45:58Z
    date available2026-08-23T07:45:58Z
    date copyright2026/04/01
    date issued2026
    identifier issn1932-6181
    identifier othermed-25-1083.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4315570
    description abstractAbstract. This study aims to introduce a wireless, remotely triggered biopsy capsule based on shape memory alloy (SMA). This capsule utilizes the negative pressure suction principle to achieve liquid sampling within the digestive tract. The proposed capsule is designed with an SMA spring-driven sampling trigger mechanism. An external magnetic field directs the capsule to the sampling area and aligns the sampling port with the sampling liquid. A Bluetooth signal is then sent and energizes the SMA spring, causing deformation and generating negative pressure for liquid sampling. The capsule prototype has dimensions of ø 15.0 mm × 30.0 mm. Both simulation and in vitro experiments have been carried out. Results showed that the proposed capsule has a rapid response time of 1.4 s and can sample liquids with viscosities ranging from 0.7 mPa·s to 100 mPa·s with the maximum sampling volume around 0.2 mL. Furthermore, the capsule's sealing performance has also been verified. The experimental results indicate that the proposed capsule can achieve Bluetooth remote-triggered liquid sampling in the digestive tract and satisfy the clinical requirements.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleStructure Design and Experimental Study on a Shape Memory Alloy Based Negative Pressure Sampling Capsule
    typeJournal Paper
    journal volume20
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4070417
    treeJournal of Medical Devices:;2026:;volume( 020 ):;issue:002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian