Development of a Noninvasive Device to Predict Postoperative Ileus Using Acoustic BiomarkersSource: Journal of Medical Devices:;2023:;volume( 017 ):;issue: 002::page 21004-1Author:Hrabe, Jennifer E.
,
Gribovskaja-Rupp, Irena
,
Guyton, Kristi
,
Kapadia, Muneera R.
,
Cromwell, John W.
DOI: 10.1115/1.4056832Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: For patients undergoing gastrointestinal surgery, postoperative ileus (POI) is difficult to predict and occurs at rates up to 30%. We hypothesized that discrete gastrointestinal acoustic biomarkers correlating to POI development may be present in early postoperative hours. To test this hypothesis, we conducted a two-phase investigation to record and analyze intestinal sounds, then to prospectively test the feasibility of predicting POI using a noninvasive device trained with an algorithm developed from the Phase 1 results. In Phase 1, a library of intestinal sounds from normal volunteer subjects and patients undergoing intestinal resection surgery were recorded. An acoustic biomarker most correlated with POI was identified and used to develop a predictive algorithm. In Phase 2, an autonomous, wearable device possessing listening and signal processing capability was developed and prospectively tested for prediction of POI in patients undergoing bowel resection surgery. In Phase 1, 30 patients undergoing intestinal resection surgery were studied, 9 of whom developed POI, from which an acoustic biomarker was identified and analyzed. In Phase 2, prospective prediction of POI in 75 enrolled subjects undergoing bowel resection surgery was assessed. POI was correctly predicted in 12 of 14 subjects who developed POI; prediction that POI would not occur was correct in 39 of 51 subjects. Sensitivity and specificity were 85.7% and 63.9%, respectively. Negative predictive value and accuracy were 95.1%, and 68%, respectively. Our study demonstrates feasibility of POI prediction based on gastrointestinal sounds using a noninvasive device. This device may help risk stratify patients likely to develop POI.
|
Collections
Show full item record
| contributor author | Hrabe, Jennifer E. | |
| contributor author | Gribovskaja-Rupp, Irena | |
| contributor author | Guyton, Kristi | |
| contributor author | Kapadia, Muneera R. | |
| contributor author | Cromwell, John W. | |
| date accessioned | 2023-08-16T18:44:58Z | |
| date available | 2023-08-16T18:44:58Z | |
| date copyright | 4/17/2023 12:00:00 AM | |
| date issued | 2023 | |
| identifier issn | 1932-6181 | |
| identifier other | med_017_02_021004.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4292429 | |
| description abstract | For patients undergoing gastrointestinal surgery, postoperative ileus (POI) is difficult to predict and occurs at rates up to 30%. We hypothesized that discrete gastrointestinal acoustic biomarkers correlating to POI development may be present in early postoperative hours. To test this hypothesis, we conducted a two-phase investigation to record and analyze intestinal sounds, then to prospectively test the feasibility of predicting POI using a noninvasive device trained with an algorithm developed from the Phase 1 results. In Phase 1, a library of intestinal sounds from normal volunteer subjects and patients undergoing intestinal resection surgery were recorded. An acoustic biomarker most correlated with POI was identified and used to develop a predictive algorithm. In Phase 2, an autonomous, wearable device possessing listening and signal processing capability was developed and prospectively tested for prediction of POI in patients undergoing bowel resection surgery. In Phase 1, 30 patients undergoing intestinal resection surgery were studied, 9 of whom developed POI, from which an acoustic biomarker was identified and analyzed. In Phase 2, prospective prediction of POI in 75 enrolled subjects undergoing bowel resection surgery was assessed. POI was correctly predicted in 12 of 14 subjects who developed POI; prediction that POI would not occur was correct in 39 of 51 subjects. Sensitivity and specificity were 85.7% and 63.9%, respectively. Negative predictive value and accuracy were 95.1%, and 68%, respectively. Our study demonstrates feasibility of POI prediction based on gastrointestinal sounds using a noninvasive device. This device may help risk stratify patients likely to develop POI. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Development of a Noninvasive Device to Predict Postoperative Ileus Using Acoustic Biomarkers | |
| type | Journal Paper | |
| journal volume | 17 | |
| journal issue | 2 | |
| journal title | Journal of Medical Devices | |
| identifier doi | 10.1115/1.4056832 | |
| journal fristpage | 21004-1 | |
| journal lastpage | 21004-7 | |
| page | 7 | |
| tree | Journal of Medical Devices:;2023:;volume( 017 ):;issue: 002 | |
| contenttype | Fulltext |