    Modification of a Simplified Hip Joint Simulator Into an ISO 14242-1 Compliant Design and a Comparison of Wear Test Results

    Source: Journal of Tribology:;2021:;volume( 144 ):;issue: 005::page 51703-1
    Author:
    Saikko, Vesa
    ,
    Morad, Omar
    ,
    Viitala, Raine
    DOI: 10.1115/1.4052942
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The ISO 14242-1 standard for hip joint simulator wear testing specifies a set of test conditions for the simulation of normal level walking in optimal conditions. Since some of the established simulators, such as the 12-station HUT-4, are not ISO 14242-1 compliant, the present study was carried out to answer the following question. Does wear produced in ISO 14242-1 conditions differ from that obtained earlier with the simplified HUT-4 hip joint simulator for similar specimens in normal level walking, optimal conditions? The established HUT-4 hip joint simulator was made ISO compliant by an implementation of a number of modifications. One of the modifications was the design and construction of a novel servo-electric load actuator with a proven dynamic response. The other modifications were related to the Euler sequence of motions, acetabular abduction angle, enclosure of the lubricant chamber, and temperature control. A 5 million-cycle wear test with thin, large-diameter VEXLPE liners resulted in a wear rate close to that obtained earlier with the HUT-4. The burnished bearing surface in both tests was in agreement with clinical retrieval studies. It appeared that a more simple, inexpensive hip joint simulator can reproduce clinical wear mechanisms. However, the simulator must meet certain basic requirements. The most important of these is likely to be the correct type of multidirectional relative motion, for which biaxial motion is sufficient. It is emphasized that the present study was not intended to show a similarity in wear produced by the ISO 14242-1 and HUT-4 wear test systems.
    • Download: (893.5Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Modification of a Simplified Hip Joint Simulator Into an ISO 14242-1 Compliant Design and a Comparison of Wear Test Results

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4284298
    Collections
    • Journal of Tribology

    Show full item record

    contributor authorSaikko, Vesa
    contributor authorMorad, Omar
    contributor authorViitala, Raine
    date accessioned2022-05-08T08:45:17Z
    date available2022-05-08T08:45:17Z
    date copyright11/25/2021 12:00:00 AM
    date issued2021
    identifier issn0742-4787
    identifier othertrib_144_5_051703.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4284298
    description abstractThe ISO 14242-1 standard for hip joint simulator wear testing specifies a set of test conditions for the simulation of normal level walking in optimal conditions. Since some of the established simulators, such as the 12-station HUT-4, are not ISO 14242-1 compliant, the present study was carried out to answer the following question. Does wear produced in ISO 14242-1 conditions differ from that obtained earlier with the simplified HUT-4 hip joint simulator for similar specimens in normal level walking, optimal conditions? The established HUT-4 hip joint simulator was made ISO compliant by an implementation of a number of modifications. One of the modifications was the design and construction of a novel servo-electric load actuator with a proven dynamic response. The other modifications were related to the Euler sequence of motions, acetabular abduction angle, enclosure of the lubricant chamber, and temperature control. A 5 million-cycle wear test with thin, large-diameter VEXLPE liners resulted in a wear rate close to that obtained earlier with the HUT-4. The burnished bearing surface in both tests was in agreement with clinical retrieval studies. It appeared that a more simple, inexpensive hip joint simulator can reproduce clinical wear mechanisms. However, the simulator must meet certain basic requirements. The most important of these is likely to be the correct type of multidirectional relative motion, for which biaxial motion is sufficient. It is emphasized that the present study was not intended to show a similarity in wear produced by the ISO 14242-1 and HUT-4 wear test systems.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleModification of a Simplified Hip Joint Simulator Into an ISO 14242-1 Compliant Design and a Comparison of Wear Test Results
    typeJournal Paper
    journal volume144
    journal issue5
    journal titleJournal of Tribology
    identifier doi10.1115/1.4052942
    journal fristpage51703-1
    journal lastpage51703-8
    page8
    treeJournal of Tribology:;2021:;volume( 144 ):;issue: 005
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian