    A New Test Method for In Vitro Evaluation of Pedicle Screw Loosening Potential

    Source: Journal of Medical Devices:;2021:;volume( 015 ):;issue: 004::page 041008-1
    Author:
    Pinto, Otávio Teixeira
    ,
    Laforce, Isabel Flesch
    ,
    Badra, Dimitra Vlaskou
    ,
    de Mello Roesler, Carlos Rodrigo
    DOI: 10.1115/1.4052517
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: This paper proposes a new test method based on the toggle effect under transverse loads (cranial–caudal) to investigate the loosening potential of pedicle screw designs. A three-step in vitro testing procedure was developed to mimic the loosening mechanism of pedicle screws. First, the pedicle screw of a certain design is inserted into a bone substitute model specifically designed for the test. Second, a controlled cyclic cranial–caudal loading is applied transversally to the longitudinal axis of the screw for three ascendent load levels (staircase) by a predetermined number of load cycles. Lastly, each pedicle screw is adjusted and subjected to axial pull-out quasi-static testing. The results are used to calculate a loosening index that, together with statistical analysis, indicates the potential for loosening of the specific design evaluated. The proposed testing method effectively provides a simulated environment to evaluate the loosening potential of pedicle screw designs. The proposed loosening index calculation may be used to compare different pedicle screw designs. The proposed methodology was verified as a valuable tool to investigate the influence of the cranial–caudal loads on pedicle screw behavior. It offers a new alternative for use in preclinical studies on the loosening potential of pedicle screw designs.
    • Download: (1.163Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      A New Test Method for In Vitro Evaluation of Pedicle Screw Loosening Potential

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4278748
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorPinto, Otávio Teixeira
    contributor authorLaforce, Isabel Flesch
    contributor authorBadra, Dimitra Vlaskou
    contributor authorde Mello Roesler, Carlos Rodrigo
    date accessioned2022-02-06T05:46:54Z
    date available2022-02-06T05:46:54Z
    date copyright10/18/2021 12:00:00 AM
    date issued2021
    identifier issn1932-6181
    identifier othermed_015_04_041008.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4278748
    description abstractThis paper proposes a new test method based on the toggle effect under transverse loads (cranial–caudal) to investigate the loosening potential of pedicle screw designs. A three-step in vitro testing procedure was developed to mimic the loosening mechanism of pedicle screws. First, the pedicle screw of a certain design is inserted into a bone substitute model specifically designed for the test. Second, a controlled cyclic cranial–caudal loading is applied transversally to the longitudinal axis of the screw for three ascendent load levels (staircase) by a predetermined number of load cycles. Lastly, each pedicle screw is adjusted and subjected to axial pull-out quasi-static testing. The results are used to calculate a loosening index that, together with statistical analysis, indicates the potential for loosening of the specific design evaluated. The proposed testing method effectively provides a simulated environment to evaluate the loosening potential of pedicle screw designs. The proposed loosening index calculation may be used to compare different pedicle screw designs. The proposed methodology was verified as a valuable tool to investigate the influence of the cranial–caudal loads on pedicle screw behavior. It offers a new alternative for use in preclinical studies on the loosening potential of pedicle screw designs.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA New Test Method for In Vitro Evaluation of Pedicle Screw Loosening Potential
    typeJournal Paper
    journal volume15
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4052517
    journal fristpage041008-1
    journal lastpage041008-6
    page6
    treeJournal of Medical Devices:;2021:;volume( 015 ):;issue: 004
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian