Show simple item record

contributor authorChiang, Luciano E.
contributor authorCastro, Felipe A.
date accessioned2022-02-06T05:46:27Z
date available2022-02-06T05:46:27Z
date copyright6/25/2021 12:00:00 AM
date issued2021
identifier issn1932-6181
identifier othermed_015_03_031008.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4278729
description abstractIn this article, we present a clinically validated invasive emergency mechanical ventilator developed in Chile called VEMERS UC. It has been clinically tested and validated in intubated Covid-19 patients. Once the pandemic hit Chilean soil in March 2020, it was quite clear that the number of mechanical ventilators available would not be enough. As in other parts of the world many initiatives sprung, most of them naively simple. Chilean medical societies joined engineering specialists and agreed early on in an organized and regulated open process for validating emergency mechanical ventilators, thus allowing for rapid development but with the required functionality, reliability, and safety features. VEMERS UC was one of a few that completed successfully all stages of the validating process, the final test being on five critically ill intubated Covid-19 patients for eight hours each. VEMERS UC is based on an electropneumatic circuit architecture, and its components are all low cost, off-the-shelf pneumatic, and electronic products easily obtained in industrial markets. It works in continuous mandatory volume control mode. The novel technical features of VEMERS UC are discussed here as well as the results obtained in each stage of the validating process. The validating process carried out in Chile is noteworthy by itself, and it could be used as an example in other developing countries. Furthermore, VEMERS UC can be used as a guiding design reference in other countries as well, since this design has already been thoroughly tested in human patients and has proven to work successfully.
publisherThe American Society of Mechanical Engineers (ASME)
titleVEMERS UC: A Clinically Validated Emergency Mechanical Ventilator for COVID-19 and Postpandemic Use in Low Resource Communities
typeJournal Paper
journal volume15
journal issue3
journal titleJournal of Medical Devices
identifier doi10.1115/1.4051246
journal fristpage031008-1
journal lastpage031008-12
page12
treeJournal of Medical Devices:;2021:;volume( 015 ):;issue: 003
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record