    Generic Implant Positioning Technology Based on Hole Projections in X-Ray Images

    Source: Journal of Medical Devices:;2021:;volume( 015 ):;issue: 002::page 025002-1
    Author:
    Windolf, Markus
    ,
    Richards, Robert Geoff
    DOI: 10.1115/1.4049979
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Implant placement plays a key role in trauma and orthopedics. In this paper, a generic technological concept for implant positioning assistance is outlined. The system utilizes conventional radiographic devices for imaging and tracking and embeds into surgical workflows without the need for complex navigation equipment. It is based on feature extraction from cylindrical hole-projections in X-ray images for determining spatial alignment of implant and anatomy. Basic performance of a prototype system was experimentally verified in terms of tracking accuracy and robustness under varying conditions. In a second step, the system was developed into a set of application modules, each serving a pressing clinical need: Plating of the proximal humerus, cephalic nail and dynamic hip-screw placement, general anatomic plating, distal nail interlocking with adjustment of femoral anteversion and corrective osteotomies. Module prototypes were tested according to their degree of maturity from feasibility assessment in wet-labs to clinical handling tests. Orientation tracking of reference objects yielded an accuracy and precision of 0.1±0.71 deg (mean±standard deviation) with a maximum error of 4.68 deg at unfavorable conditions. This base-performance translated, e.g., into a precision of ±1.2 mm (standard deviation) screw-tip to joint distance at proximal humerus plating, or into a precision of lag screw positioning in the femoral head of ±0.6 mm in craniocaudal and ±1.6 mm in anterioposterior direction. The concept revealed strong potential to improve surgical outcomes in a broad range of orthopedic applications due to its generic and simplistic nature. Comprehensive validation activities must follow for clinical introduction.
    • Download: (3.000Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Generic Implant Positioning Technology Based on Hole Projections in X-Ray Images

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4276437
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorWindolf, Markus
    contributor authorRichards, Robert Geoff
    date accessioned2022-02-05T21:50:15Z
    date available2022-02-05T21:50:15Z
    date copyright3/23/2021 12:00:00 AM
    date issued2021
    identifier issn1932-6181
    identifier othermed_015_02_025002.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4276437
    description abstractImplant placement plays a key role in trauma and orthopedics. In this paper, a generic technological concept for implant positioning assistance is outlined. The system utilizes conventional radiographic devices for imaging and tracking and embeds into surgical workflows without the need for complex navigation equipment. It is based on feature extraction from cylindrical hole-projections in X-ray images for determining spatial alignment of implant and anatomy. Basic performance of a prototype system was experimentally verified in terms of tracking accuracy and robustness under varying conditions. In a second step, the system was developed into a set of application modules, each serving a pressing clinical need: Plating of the proximal humerus, cephalic nail and dynamic hip-screw placement, general anatomic plating, distal nail interlocking with adjustment of femoral anteversion and corrective osteotomies. Module prototypes were tested according to their degree of maturity from feasibility assessment in wet-labs to clinical handling tests. Orientation tracking of reference objects yielded an accuracy and precision of 0.1±0.71 deg (mean±standard deviation) with a maximum error of 4.68 deg at unfavorable conditions. This base-performance translated, e.g., into a precision of ±1.2 mm (standard deviation) screw-tip to joint distance at proximal humerus plating, or into a precision of lag screw positioning in the femoral head of ±0.6 mm in craniocaudal and ±1.6 mm in anterioposterior direction. The concept revealed strong potential to improve surgical outcomes in a broad range of orthopedic applications due to its generic and simplistic nature. Comprehensive validation activities must follow for clinical introduction.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleGeneric Implant Positioning Technology Based on Hole Projections in X-Ray Images
    typeJournal Paper
    journal volume15
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4049979
    journal fristpage025002-1
    journal lastpage025002-11
    page11
    treeJournal of Medical Devices:;2021:;volume( 015 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian