    Design and Proof of Principle of a Novel Drip Chamber

    Source: Journal of Medical Devices:;2020:;volume( 014 ):;issue: 002
    Author:
    de Ridder, Else F.
    ,
    Hoeboer, Max D. B.
    ,
    van Dam, Anne F. D. M.
    ,
    van den Bosch, Job
    DOI: 10.1115/1.4045690
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: More than a million times every year, an ambulance is deployed in the Netherlands. Of these instances, 87% of the patients will be administered intravenous (IV) fluids. In the current situation, the IV bag always needs to be held above the patient to function properly. This action requires an extra pair of hands, which is very inefficient and can result in the loss of precious time that could otherwise be directed at the patient. Besides needing an extra pair of hands, there is also a chance of air bubbles entering the patient through the drip feed. The novel drip chamber design proposed in this paper aims to solve these problems, as it enables the IV bag to be placed in any orientation and in any manner relative to the patient with a negligible number of air bubbles entering the drip feed. This novel drip chamber was tested in an experimental setting in different orientations and at different flowrates. From the test results, it can be concluded that at a clinical relevant flowrate, a negligible number of air bubbles were present within the IV system. Because of the ease of use of the novel drip chamber and the fact that it is fool-proof, cost-efficient, and shows promising test results, future research on several aspects could make this product a promising addition to health care.
    • Download: (471.7Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Design and Proof of Principle of a Novel Drip Chamber

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4273738
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorde Ridder, Else F.
    contributor authorHoeboer, Max D. B.
    contributor authorvan Dam, Anne F. D. M.
    contributor authorvan den Bosch, Job
    date accessioned2022-02-04T14:28:48Z
    date available2022-02-04T14:28:48Z
    date copyright2020/02/18/
    date issued2020
    identifier issn1932-6181
    identifier othermed_014_02_025004.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4273738
    description abstractMore than a million times every year, an ambulance is deployed in the Netherlands. Of these instances, 87% of the patients will be administered intravenous (IV) fluids. In the current situation, the IV bag always needs to be held above the patient to function properly. This action requires an extra pair of hands, which is very inefficient and can result in the loss of precious time that could otherwise be directed at the patient. Besides needing an extra pair of hands, there is also a chance of air bubbles entering the patient through the drip feed. The novel drip chamber design proposed in this paper aims to solve these problems, as it enables the IV bag to be placed in any orientation and in any manner relative to the patient with a negligible number of air bubbles entering the drip feed. This novel drip chamber was tested in an experimental setting in different orientations and at different flowrates. From the test results, it can be concluded that at a clinical relevant flowrate, a negligible number of air bubbles were present within the IV system. Because of the ease of use of the novel drip chamber and the fact that it is fool-proof, cost-efficient, and shows promising test results, future research on several aspects could make this product a promising addition to health care.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign and Proof of Principle of a Novel Drip Chamber
    typeJournal Paper
    journal volume14
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4045690
    page25004
    treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian