    Archive

    Avoiding the Alveolar Nerve Via a Real-Time Impedance Analysis: A Novel Method to Improve Implant Surgery Safety

    Source: Journal of Medical Devices:;2020:;volume( 014 ):;issue: 002
    Author:
    Schiffers, Georg
    ,
    Arnold, Julian
    ,
    Bartella, Alexander
    ,
    Mähönen, Petri
    ,
    Lethaus, Bernd
    DOI: 10.1115/1.4046212
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Safe lower jaw implantations require precautions to avoid damaging the alveolar nerve. The prevailing methods are preoperative. In this study, we propose a novel approach to measure the distance between a pilot-drill and the alveolar nerve by employing high-frequency impedance measurements. The objective is to provide in vivo real-time information as an early warning of the proximity of the alveolar nerve. The method is examined and tested on animal samples. The impedance measurements were performed using a high-frequency network analyzer. Overall 40 pilot drillings were distributed over five sheep mandibles, with four on each side. Drillings were performed in three steps: inside the cortical layer, inside the spongiosa, and well inside the nerve canal. The inductance measurements were performed with a connected pilot drill, followed by an immediate 3D cone-beam computed tomography (CT) to measure the distance between the tip of the drill and the nerve canal. The measurements show that impedance information is a reliable indicator for proximity of the drill to the nerve. We observe a general trend of decreasing inductance as the drill approaches the nerve and find that at very high frequencies one can detect the closeness to the nerve from characteristic ratios of impedance at nearby frequencies. We report also that using phase information increases the reliability of this method. The findings provide a solid proof of concept for the proposed method. While the results are promising at this stage, the applicability for in vivo conditions requires further studies.
    • Download: (1.985Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Avoiding the Alveolar Nerve Via a Real-Time Impedance Analysis: A Novel Method to Improve Implant Surgery Safety

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4273441
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorSchiffers, Georg
    contributor authorArnold, Julian
    contributor authorBartella, Alexander
    contributor authorMähönen, Petri
    contributor authorLethaus, Bernd
    date accessioned2022-02-04T14:19:45Z
    date available2022-02-04T14:19:45Z
    date copyright2020/02/18/
    date issued2020
    identifier issn1932-6181
    identifier othermed_014_02_021003.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4273441
    description abstractSafe lower jaw implantations require precautions to avoid damaging the alveolar nerve. The prevailing methods are preoperative. In this study, we propose a novel approach to measure the distance between a pilot-drill and the alveolar nerve by employing high-frequency impedance measurements. The objective is to provide in vivo real-time information as an early warning of the proximity of the alveolar nerve. The method is examined and tested on animal samples. The impedance measurements were performed using a high-frequency network analyzer. Overall 40 pilot drillings were distributed over five sheep mandibles, with four on each side. Drillings were performed in three steps: inside the cortical layer, inside the spongiosa, and well inside the nerve canal. The inductance measurements were performed with a connected pilot drill, followed by an immediate 3D cone-beam computed tomography (CT) to measure the distance between the tip of the drill and the nerve canal. The measurements show that impedance information is a reliable indicator for proximity of the drill to the nerve. We observe a general trend of decreasing inductance as the drill approaches the nerve and find that at very high frequencies one can detect the closeness to the nerve from characteristic ratios of impedance at nearby frequencies. We report also that using phase information increases the reliability of this method. The findings provide a solid proof of concept for the proposed method. While the results are promising at this stage, the applicability for in vivo conditions requires further studies.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleAvoiding the Alveolar Nerve Via a Real-Time Impedance Analysis: A Novel Method to Improve Implant Surgery Safety
    typeJournal Paper
    journal volume14
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4046212
    page21003
    treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian