Design of a Metered-Dose Inhaler Actuated by Shape Memory AlloySource: Journal of Engineering and Science in Medical Diagnostics and Therapy:;2019:;volume( 002 ):;issue: 003::page 31003Author:Utter, Brent
DOI: 10.1115/1.4043403Publisher: American Society of Mechanical Engineers (ASME)
Abstract: The global prevalence of asthma and chronic obstructive pulmonary disease (COPD) is on the order of hundreds of millions of individuals. The most common treatment approach is to take a self-administered inhaled medication. This study focuses on pressurized metered-dose inhalers (MDIs) where, unfortunately, rates of mishandling and misuse are extremely high and lead to improper treatment. One significant challenge results from the timing miscoordination of the medicine dispersion and inhalation breath. To address this, this study demonstrates the feasibility of automating the timing of the medicine dispersion by integrating a shape memory alloy (SMA) actuator and a differential pressure sensor into the casing of a traditional MDI. The approach is to measure the vacuum pressure created by an inspiratory breath, evaluate criteria indicating an acceptable breath, and if those criteria are met, heat the SMA actuator to depress the cartridge and disperse medicine. To meet actuation requirements and reliably depress the inhaler cartridge, two concepts for configuring an SMA wire were designed and compared with respect to complexity, actuation timing, and energy consumption. The proposed concept was able to disperse medicine in 263 ms, averaged over 100 actuations on a single battery charge, facilitating the early dispersion of medicine during an inhalation breath. By describing the design process of an SMA-actuated MDI that does not result in a significant increase of its weight or size, this study provides a practical technological approach for reducing the improper treatment of asthma and COPD due to timing miscoordination.
|
Show full item record
| contributor author | Utter, Brent | |
| date accessioned | 2019-09-18T09:07:39Z | |
| date available | 2019-09-18T09:07:39Z | |
| date copyright | 5/8/2019 12:00:00 AM | |
| date issued | 2019 | |
| identifier issn | 2572-7958 | |
| identifier other | jesmdt_002_03_031003 | |
| identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4259176 | |
| description abstract | The global prevalence of asthma and chronic obstructive pulmonary disease (COPD) is on the order of hundreds of millions of individuals. The most common treatment approach is to take a self-administered inhaled medication. This study focuses on pressurized metered-dose inhalers (MDIs) where, unfortunately, rates of mishandling and misuse are extremely high and lead to improper treatment. One significant challenge results from the timing miscoordination of the medicine dispersion and inhalation breath. To address this, this study demonstrates the feasibility of automating the timing of the medicine dispersion by integrating a shape memory alloy (SMA) actuator and a differential pressure sensor into the casing of a traditional MDI. The approach is to measure the vacuum pressure created by an inspiratory breath, evaluate criteria indicating an acceptable breath, and if those criteria are met, heat the SMA actuator to depress the cartridge and disperse medicine. To meet actuation requirements and reliably depress the inhaler cartridge, two concepts for configuring an SMA wire were designed and compared with respect to complexity, actuation timing, and energy consumption. The proposed concept was able to disperse medicine in 263 ms, averaged over 100 actuations on a single battery charge, facilitating the early dispersion of medicine during an inhalation breath. By describing the design process of an SMA-actuated MDI that does not result in a significant increase of its weight or size, this study provides a practical technological approach for reducing the improper treatment of asthma and COPD due to timing miscoordination. | |
| publisher | American Society of Mechanical Engineers (ASME) | |
| title | Design of a Metered-Dose Inhaler Actuated by Shape Memory Alloy | |
| type | Journal Paper | |
| journal volume | 2 | |
| journal issue | 3 | |
| journal title | Journal of Engineering and Science in Medical Diagnostics and Therapy | |
| identifier doi | 10.1115/1.4043403 | |
| journal fristpage | 31003 | |
| journal lastpage | 031003-8 | |
| tree | Journal of Engineering and Science in Medical Diagnostics and Therapy:;2019:;volume( 002 ):;issue: 003 | |
| contenttype | Fulltext |