    An Electronic Force Sensor for Medical Jet Injection

    Source: Journal of Medical Devices:;2019:;volume( 013 ):;issue: 002::page 21007
    Author:
    Demas, Nickolas P.
    ,
    Hunter, Ian W.
    DOI: 10.1115/1.4043196
    Publisher: American Society of Mechanical Engineers (ASME)
    Abstract: In medical jet injection, a narrow fluid drug stream is propelled at high velocity into skin without a needle. Previous studies have shown that the volume delivered is highly dependent on a number of factors. This paper details the development of an electronic force sensor for medical jet injection and shows that the normal contact force exerted on the tissue by the nozzle is an additional factor affecting volume delivered. Using this sensor, we measure the forces at the nozzle tip in the normal direction with a sensitivity of 18 μN, calibrated over a range from 1 N to 8 N with a mean absolute error of 8 mN, and a maximum overload of 300 N. We further measure forces at the nozzle tip in the lateral direction with a sensitivity of 8 μN, calibrated over a range from 0.1 N to 7 N, with a mean absolute error of 101 mN for lateral contact force magnitude and 1.60 deg for lateral contact force direction. Experimental validation confirms that the force sensor does not adversely affect the accuracy and precision of ejected volume from the jet injector. We use this setup to examine the effect of normal contact force on volume delivered into postmortem porcine tissue. Experimental results demonstrate that volume delivered with normal contact force between 4 N and 8 N is significantly more accurate and precise compared to volume delivered with normal contact force between 0 N and 3.9 N.
    • Download: (4.245Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      An Electronic Force Sensor for Medical Jet Injection

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4258980
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorDemas, Nickolas P.
    contributor authorHunter, Ian W.
    date accessioned2019-09-18T09:06:39Z
    date available2019-09-18T09:06:39Z
    date copyright4/11/2019 12:00:00 AM
    date issued2019
    identifier issn1932-6181
    identifier othermed_013_02_021007
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4258980
    description abstractIn medical jet injection, a narrow fluid drug stream is propelled at high velocity into skin without a needle. Previous studies have shown that the volume delivered is highly dependent on a number of factors. This paper details the development of an electronic force sensor for medical jet injection and shows that the normal contact force exerted on the tissue by the nozzle is an additional factor affecting volume delivered. Using this sensor, we measure the forces at the nozzle tip in the normal direction with a sensitivity of 18 μN, calibrated over a range from 1 N to 8 N with a mean absolute error of 8 mN, and a maximum overload of 300 N. We further measure forces at the nozzle tip in the lateral direction with a sensitivity of 8 μN, calibrated over a range from 0.1 N to 7 N, with a mean absolute error of 101 mN for lateral contact force magnitude and 1.60 deg for lateral contact force direction. Experimental validation confirms that the force sensor does not adversely affect the accuracy and precision of ejected volume from the jet injector. We use this setup to examine the effect of normal contact force on volume delivered into postmortem porcine tissue. Experimental results demonstrate that volume delivered with normal contact force between 4 N and 8 N is significantly more accurate and precise compared to volume delivered with normal contact force between 0 N and 3.9 N.
    publisherAmerican Society of Mechanical Engineers (ASME)
    titleAn Electronic Force Sensor for Medical Jet Injection
    typeJournal Paper
    journal volume13
    journal issue2
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4043196
    journal fristpage21007
    journal lastpage021007-8
    treeJournal of Medical Devices:;2019:;volume( 013 ):;issue: 002
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian