YaBeSH Engineering and Technology Library

    • Journals
    • PaperQuest
    • YSE Standards
    • YaBeSH
    • Login
    View Item 
    •   YE&T Library
    • ASME
    • Journal of Medical Devices
    • View Item
    •   YE&T Library
    • ASME
    • Journal of Medical Devices
    • View Item
    • All Fields
    • Source Title
    • Year
    • Publisher
    • Title
    • Subject
    • Author
    • DOI
    • ISBN
    Advanced Search
    JavaScript is disabled for your browser. Some features of this site may not work without it.

    Archive

    A Novel Modular Tonometry-Based Device to Measure Pulse Pressure Waveforms in Radial Artery

    Source: Journal of Medical Devices:;2018:;volume( 012 ):;issue: 001::page 11011
    Author:
    Choudhury, Mohammad Ikbal
    ,
    Singh, Pranjal
    ,
    Juneja, Rajneesh
    ,
    Tuli, Suneet
    ,
    Deepak, K. K.
    ,
    Prasad, Anamika
    ,
    Roy, Sitikantha
    DOI: 10.1115/1.4039010
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: The paper presents the development of a new device for measuring continuous pulse pressure waveforms (PPW) from the radial artery via applanation tonometry. The development focuses on improved accuracy, open and affordable design using off-the-shelf components, and greater user control in setting operational and calibration parameters to address user variability. The device design parameters are optimized through a tissue device interaction study based on a computational model. The design incorporates modular components and includes a sensor module for arterial flattening and pressure pick-up, a differential screw mechanism and a related algorithm for controlled stepwise motion and data collection during flattening, and a brace for wrist-flexion adjustment. Maximum pulse amplitude (PA) was used as an indicator of the optimum level of arterial flattening for recording the PPW. The PPW was observed to distort due to changes in parameters like gel-head placement, hold-down pressure (HDP), and wrist extension. The pressure waveforms collected using the device were validated using limited data against established products and showed good correlation within ±1.96 standard deviation of the mean difference in a Bland–Altman plot. This paper thus details the development of a simple and validated mechanical design to measure PPW using arterial tonometry.
    • Download: (3.592Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      A Novel Modular Tonometry-Based Device to Measure Pulse Pressure Waveforms in Radial Artery

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4252441
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorChoudhury, Mohammad Ikbal
    contributor authorSingh, Pranjal
    contributor authorJuneja, Rajneesh
    contributor authorTuli, Suneet
    contributor authorDeepak, K. K.
    contributor authorPrasad, Anamika
    contributor authorRoy, Sitikantha
    date accessioned2019-02-28T11:04:45Z
    date available2019-02-28T11:04:45Z
    date copyright1/31/2018 12:00:00 AM
    date issued2018
    identifier issn1932-6181
    identifier othermed_012_01_011011.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252441
    description abstractThe paper presents the development of a new device for measuring continuous pulse pressure waveforms (PPW) from the radial artery via applanation tonometry. The development focuses on improved accuracy, open and affordable design using off-the-shelf components, and greater user control in setting operational and calibration parameters to address user variability. The device design parameters are optimized through a tissue device interaction study based on a computational model. The design incorporates modular components and includes a sensor module for arterial flattening and pressure pick-up, a differential screw mechanism and a related algorithm for controlled stepwise motion and data collection during flattening, and a brace for wrist-flexion adjustment. Maximum pulse amplitude (PA) was used as an indicator of the optimum level of arterial flattening for recording the PPW. The PPW was observed to distort due to changes in parameters like gel-head placement, hold-down pressure (HDP), and wrist extension. The pressure waveforms collected using the device were validated using limited data against established products and showed good correlation within ±1.96 standard deviation of the mean difference in a Bland–Altman plot. This paper thus details the development of a simple and validated mechanical design to measure PPW using arterial tonometry.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleA Novel Modular Tonometry-Based Device to Measure Pulse Pressure Waveforms in Radial Artery
    typeJournal Paper
    journal volume12
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4039010
    journal fristpage11011
    journal lastpage011011-10
    treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian