    An Inertia-Deformability Hybrid Circulating Tumor Cell Chip: Design, Clinical Test, and Numerical Analysis

    Source: Journal of Medical Devices:;2018:;volume( 012 ):;issue: 004::page 41004
    Author:
    Chen, Hongmei
    ,
    Zhang, Zhifeng
    DOI: 10.1115/1.4040986
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Detection and capture of circulating tumor cells (CTCs) with microfluidic chips hold significance in cancer prognosis, diagnosis, and anti-cancer treatment. The counting of CTCs provides potential tools to evaluate cancer stages as well as treatment progress. However, facing the challenge of rareness in blood, the precise enumeration of CTCs is challenging. In the present research, we designed an inertial-deformability hybrid microfluidic chip using a long spiral channel with trapezoid-circular pillars and a capture zone. To clinically validate the device, the microfluidic chip has been tested for the whole blood and lysed blood with a small number of CTCs (colorectal and nonsmall-cell lung cancer) spiked in. The capture efficiency reaches over 90% for three types of cancer cell lines at the flow rate of 1.5 mL/h. Following numerical modeling was conducted to explain the working principle and working condition (Reynolds number below 10 and Dean number around 1). This design extended the effective capture length, improved the capture efficiency, and made the CTC enumeration much easier. We believe that this hybrid chip is promising clinically in the CTCs enumeration, evaluation of cancer therapy, and pharmacological responses.
    • Download: (1.662Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      An Inertia-Deformability Hybrid Circulating Tumor Cell Chip: Design, Clinical Test, and Numerical Analysis

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4252440
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorChen, Hongmei
    contributor authorZhang, Zhifeng
    date accessioned2019-02-28T11:04:44Z
    date available2019-02-28T11:04:44Z
    date copyright9/21/2018 12:00:00 AM
    date issued2018
    identifier issn1932-6181
    identifier othermed_012_04_041004.pdf
    identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252440
    description abstractDetection and capture of circulating tumor cells (CTCs) with microfluidic chips hold significance in cancer prognosis, diagnosis, and anti-cancer treatment. The counting of CTCs provides potential tools to evaluate cancer stages as well as treatment progress. However, facing the challenge of rareness in blood, the precise enumeration of CTCs is challenging. In the present research, we designed an inertial-deformability hybrid microfluidic chip using a long spiral channel with trapezoid-circular pillars and a capture zone. To clinically validate the device, the microfluidic chip has been tested for the whole blood and lysed blood with a small number of CTCs (colorectal and nonsmall-cell lung cancer) spiked in. The capture efficiency reaches over 90% for three types of cancer cell lines at the flow rate of 1.5 mL/h. Following numerical modeling was conducted to explain the working principle and working condition (Reynolds number below 10 and Dean number around 1). This design extended the effective capture length, improved the capture efficiency, and made the CTC enumeration much easier. We believe that this hybrid chip is promising clinically in the CTCs enumeration, evaluation of cancer therapy, and pharmacological responses.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleAn Inertia-Deformability Hybrid Circulating Tumor Cell Chip: Design, Clinical Test, and Numerical Analysis
    typeJournal Paper
    journal volume12
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4040986
    journal fristpage41004
    journal lastpage041004-6
    treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 004
    contenttypeFulltext