    Design and Preliminary Testing of Novel Injection Port Contamination Barrier Devices

    Source: Journal of Medical Devices:;2017:;volume( 011 ):;issue: 003::page 34503
    Author:
    Wax, David B.
    ,
    Hill, Bryan
    DOI: 10.1115/1.4036026
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Prior studies have linked microbial contamination of intravenous (IV) ports and stopcocks with postoperative infections. Existing technologies to address contamination are not consistently utilized because of the time and effort they require. Herein, novel barrier devices were created that form a protective shell to passively prevent contact between injection sites and practitioner hands or environmental surfaces while still allowing rapid connection of a syringe for injection of medications via an opening in the shell. Prototypes were tested using a grossly contaminated environment and adenosine triphosphate (ATP)-bioluminescence assay. For eight pairs of unshielded versus shielded IV ports/stopcocks, average contamination was 4102 versus 35 RLU (p < 0.02), respectively, indicating that the devices could significantly reduce IV port/stopcock contamination.
    • Download: (827.8Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Design and Preliminary Testing of Novel Injection Port Contamination Barrier Devices

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/4235242
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorWax, David B.
    contributor authorHill, Bryan
    date accessioned2017-11-25T07:18:33Z
    date available2017-11-25T07:18:33Z
    date copyright2017/27/6
    date issued2017
    identifier issn1932-6181
    identifier othermed_011_03_034503.pdf
    identifier urihttp://138.201.223.254:8080/yetl1/handle/yetl/4235242
    description abstractPrior studies have linked microbial contamination of intravenous (IV) ports and stopcocks with postoperative infections. Existing technologies to address contamination are not consistently utilized because of the time and effort they require. Herein, novel barrier devices were created that form a protective shell to passively prevent contact between injection sites and practitioner hands or environmental surfaces while still allowing rapid connection of a syringe for injection of medications via an opening in the shell. Prototypes were tested using a grossly contaminated environment and adenosine triphosphate (ATP)-bioluminescence assay. For eight pairs of unshielded versus shielded IV ports/stopcocks, average contamination was 4102 versus 35 RLU (p < 0.02), respectively, indicating that the devices could significantly reduce IV port/stopcock contamination.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleDesign and Preliminary Testing of Novel Injection Port Contamination Barrier Devices
    typeJournal Paper
    journal volume11
    journal issue3
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4036026
    journal fristpage34503
    journal lastpage034503-2
    treeJournal of Medical Devices:;2017:;volume( 011 ):;issue: 003
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian