Device for Verifying the Patellar Cut During Knee Replacement SurgerySource: Journal of Medical Devices:;2016:;volume( 010 ):;issue: 002::page 24502Author:Rex, Erica L.
,
Illical, Emmanuel M.
,
Gaudelli, Cinzia
,
Wylant, Barry
,
Ho, Karen C. T.
,
Person, John G.
,
Anglin, Carolyn
DOI: 10.1115/1.4032870Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: Cutting the kneecap (patella) in knee replacement surgery is challenging and can lead to pain and reduced function when done incorrectly. The presented device allows the surgeon to check the threedimensional symmetry and thickness of the bone remnant before the operation is complete. Observations and measurements made on 36 resected artificial patellae and 16 resected cadaveric patellae matched well with computed tomography (CT) scans of the patellae with few exceptions; the exceptions should be addressable by changes in design and use. Average time to apply the device was 1 min.
|
Collections
Show full item record
| contributor author | Rex, Erica L. | |
| contributor author | Illical, Emmanuel M. | |
| contributor author | Gaudelli, Cinzia | |
| contributor author | Wylant, Barry | |
| contributor author | Ho, Karen C. T. | |
| contributor author | Person, John G. | |
| contributor author | Anglin, Carolyn | |
| date accessioned | 2017-05-09T01:31:43Z | |
| date available | 2017-05-09T01:31:43Z | |
| date issued | 2016 | |
| identifier issn | 1932-6181 | |
| identifier other | med_010_02_024502.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/161996 | |
| description abstract | Cutting the kneecap (patella) in knee replacement surgery is challenging and can lead to pain and reduced function when done incorrectly. The presented device allows the surgeon to check the threedimensional symmetry and thickness of the bone remnant before the operation is complete. Observations and measurements made on 36 resected artificial patellae and 16 resected cadaveric patellae matched well with computed tomography (CT) scans of the patellae with few exceptions; the exceptions should be addressable by changes in design and use. Average time to apply the device was 1 min. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Device for Verifying the Patellar Cut During Knee Replacement Surgery | |
| type | Journal Paper | |
| journal volume | 10 | |
| journal issue | 2 | |
| journal title | Journal of Medical Devices | |
| identifier doi | 10.1115/1.4032870 | |
| journal fristpage | 24502 | |
| journal lastpage | 24502 | |
| identifier eissn | 1932-619X | |
| tree | Journal of Medical Devices:;2016:;volume( 010 ):;issue: 002 | |
| contenttype | Fulltext |