    Effect of C Arm Year Model on Radiation Scatter

    Source: Journal of Medical Devices:;2015:;volume( 009 ):;issue: 001::page 11010
    Author:
    Bakr, Omar
    ,
    Morshed, Saam
    ,
    Marmor, Meir
    DOI: 10.1115/1.4029419
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: Since the 1980’s Carm fluoroscopy has been an integral part of orthopaedic trauma surgery. The advancement in Carm technology has resulted in different generations of Carms coexisting in the operating rooms. The purpose of this study was to compare the radiation scatter patterns of different generation Carms. Three generation of Carms were tested: GE OEC 9800 Plus (1999/2000), Siemens Arcadis Orbic 3D (2004), Philips BV Pulsera 2.3 (2008). Radiation scatter was measured using six realtime dosimetry badges set up on either side of the surgical table (Mizuho OSI, flattop). Distance of Carm was normalized at 20 in. and 10 in. from Image Intensifier. Each device was set to the automatic brightness control (ABC) setting. A phantom limb was irradiated for 120 s and radiation scatter was summed for both AP and lateral positions. At their typical operating room settings there was a reduction in radiation scatter using the newer generation Carms. Results for total radiation, normalized to Philips, are as follows: Philips 1 (100%), GE 2.4 (240%), and Siemens 1.4 (140%). Newer generation Carms can be expected to generate lower radiation scatter. Special care should be taken to attempt a lower dose setting, especially when utilizing older generation Carms to minimize radiation scatter to practitioner.
    • Download: (802.7Kb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Effect of C Arm Year Model on Radiation Scatter

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/159039
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorBakr, Omar
    contributor authorMorshed, Saam
    contributor authorMarmor, Meir
    date accessioned2017-05-09T01:21:35Z
    date available2017-05-09T01:21:35Z
    date issued2015
    identifier issn1932-6181
    identifier othermed_009_01_011010.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/159039
    description abstractSince the 1980’s Carm fluoroscopy has been an integral part of orthopaedic trauma surgery. The advancement in Carm technology has resulted in different generations of Carms coexisting in the operating rooms. The purpose of this study was to compare the radiation scatter patterns of different generation Carms. Three generation of Carms were tested: GE OEC 9800 Plus (1999/2000), Siemens Arcadis Orbic 3D (2004), Philips BV Pulsera 2.3 (2008). Radiation scatter was measured using six realtime dosimetry badges set up on either side of the surgical table (Mizuho OSI, flattop). Distance of Carm was normalized at 20 in. and 10 in. from Image Intensifier. Each device was set to the automatic brightness control (ABC) setting. A phantom limb was irradiated for 120 s and radiation scatter was summed for both AP and lateral positions. At their typical operating room settings there was a reduction in radiation scatter using the newer generation Carms. Results for total radiation, normalized to Philips, are as follows: Philips 1 (100%), GE 2.4 (240%), and Siemens 1.4 (140%). Newer generation Carms can be expected to generate lower radiation scatter. Special care should be taken to attempt a lower dose setting, especially when utilizing older generation Carms to minimize radiation scatter to practitioner.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleEffect of C Arm Year Model on Radiation Scatter
    typeJournal Paper
    journal volume9
    journal issue1
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4029419
    journal fristpage11010
    journal lastpage11010
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2015:;volume( 009 ):;issue: 001
    contenttypeFulltext
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian
     
    DSpace software copyright © 2002-2015  DuraSpace
    نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
    yabeshDSpacePersian