The Influence of the Aortic Root Geometry on Flow Characteristics of a Prosthetic Heart ValveSource: Journal of Biomechanical Engineering:;2015:;volume( 137 ):;issue: 005::page 51005DOI: 10.1115/1.4029747Publisher: The American Society of Mechanical Engineers (ASME)
Abstract: In this paper, performance of aortic heart valve prosthesis in different geometries of the aortic root is investigated experimentally. The objective of this investigation is to establish a set of parameters, which are associated with abnormal flow patterns due to the flow through a prosthetic heart valve implanted in the patients that had certain types of valve diseases prior to the valve replacement. Specific valve diseases were classified into two clinical categories and were correlated with the corresponding changes in aortic root geometry while keeping the aortic base diameter fixed. These categories correspond to aortic valve stenosis and aortic valve insufficiency. The control case that corresponds to the aortic root of a patient without valve disease was used as a reference. Experiments were performed at test conditions corresponding to 70 beats/min, 5.5 L/min target cardiac output, and a mean aortic pressure of 100 mmHg. By varying the aortic root geometry, while keeping the diameter of the orifice constant, it was possible to investigate corresponding changes in the levels of Reynolds shear stress and establish the possibility of platelet activation and, as a result of that, the formation of blood clots.
|
Collections
Show full item record
| contributor author | Barannyk, Oleksandr | |
| contributor author | Oshkai, Peter | |
| date accessioned | 2017-05-09T01:15:08Z | |
| date available | 2017-05-09T01:15:08Z | |
| date issued | 2015 | |
| identifier issn | 0148-0731 | |
| identifier other | bio_137_05_051005.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/157111 | |
| description abstract | In this paper, performance of aortic heart valve prosthesis in different geometries of the aortic root is investigated experimentally. The objective of this investigation is to establish a set of parameters, which are associated with abnormal flow patterns due to the flow through a prosthetic heart valve implanted in the patients that had certain types of valve diseases prior to the valve replacement. Specific valve diseases were classified into two clinical categories and were correlated with the corresponding changes in aortic root geometry while keeping the aortic base diameter fixed. These categories correspond to aortic valve stenosis and aortic valve insufficiency. The control case that corresponds to the aortic root of a patient without valve disease was used as a reference. Experiments were performed at test conditions corresponding to 70 beats/min, 5.5 L/min target cardiac output, and a mean aortic pressure of 100 mmHg. By varying the aortic root geometry, while keeping the diameter of the orifice constant, it was possible to investigate corresponding changes in the levels of Reynolds shear stress and establish the possibility of platelet activation and, as a result of that, the formation of blood clots. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | The Influence of the Aortic Root Geometry on Flow Characteristics of a Prosthetic Heart Valve | |
| type | Journal Paper | |
| journal volume | 137 | |
| journal issue | 5 | |
| journal title | Journal of Biomechanical Engineering | |
| identifier doi | 10.1115/1.4029747 | |
| journal fristpage | 51005 | |
| journal lastpage | 51005 | |
| identifier eissn | 1528-8951 | |
| tree | Journal of Biomechanical Engineering:;2015:;volume( 137 ):;issue: 005 | |
| contenttype | Fulltext |