    Mechatronic System for ABO Human Blood Typing

    Source: Journal of Medical Devices:;2014:;volume( 008 ):;issue: 004::page 44507
    Author:
    Moreira, Vأ¢nia
    ,
    Machado, Josأ©
    ,
    Carvalho, Vأ­tor
    ,
    Soares, Filomena
    ,
    Ferraz, Ana
    DOI: 10.1115/1.4028201
    Publisher: The American Society of Mechanical Engineers (ASME)
    Abstract: In medical emergency situations, when a patient needs a blood transfusion, the universal blood type O− is administered. This procedure may lead to the depletion of stock reserves of O− blood. Nowadays, there is no commercial equipment capable of determining the patient's blood type in situ, in a fast and reliable process. Human blood typing is usually performed through the manual test, which involves a macroscopic observation and interpretation of the results by an analyst. This test, despite of having a fast response time, may lead to human errors, which sometimes can be fatal to the patient. This paper presents the development of an automatic mechatronic prototype for determining human blood typing (ABO and Rh systems) through image processing techniques. The prototype design takes into account the characteristics of reliability of analysis, portability, and response time allowing the system to be used in emergency situations. The developed prototype performs blood and reagents mixture acquires the resultant image and processes the data (based on image processing techniques) to determine the sample blood type. It was tested in a laboratory, using cataloged samples of blood types, provided by the Portuguese Institute of Blood and Transplantation. Hereafter, it is expected to test and validate the prototype in clinical environments.
    • Download: (1.632Mb)
    • Show Full MetaData Hide Full MetaData
    • Get RIS
    • Item Order
    • Go To Publisher
    • Statistics

      Mechatronic System for ABO Human Blood Typing

    URI
    https://yetl.yabesh.ir/yetl1/handle/yetl/155976
    Collections
    • Journal of Medical Devices

    Show full item record

    contributor authorMoreira, Vأ¢nia
    contributor authorMachado, Josأ©
    contributor authorCarvalho, Vأ­tor
    contributor authorSoares, Filomena
    contributor authorFerraz, Ana
    date accessioned2017-05-09T01:11:24Z
    date available2017-05-09T01:11:24Z
    date issued2014
    identifier issn1932-6181
    identifier othermed_008_04_044507.pdf
    identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/155976
    description abstractIn medical emergency situations, when a patient needs a blood transfusion, the universal blood type O− is administered. This procedure may lead to the depletion of stock reserves of O− blood. Nowadays, there is no commercial equipment capable of determining the patient's blood type in situ, in a fast and reliable process. Human blood typing is usually performed through the manual test, which involves a macroscopic observation and interpretation of the results by an analyst. This test, despite of having a fast response time, may lead to human errors, which sometimes can be fatal to the patient. This paper presents the development of an automatic mechatronic prototype for determining human blood typing (ABO and Rh systems) through image processing techniques. The prototype design takes into account the characteristics of reliability of analysis, portability, and response time allowing the system to be used in emergency situations. The developed prototype performs blood and reagents mixture acquires the resultant image and processes the data (based on image processing techniques) to determine the sample blood type. It was tested in a laboratory, using cataloged samples of blood types, provided by the Portuguese Institute of Blood and Transplantation. Hereafter, it is expected to test and validate the prototype in clinical environments.
    publisherThe American Society of Mechanical Engineers (ASME)
    titleMechatronic System for ABO Human Blood Typing
    typeJournal Paper
    journal volume8
    journal issue4
    journal titleJournal of Medical Devices
    identifier doi10.1115/1.4028201
    journal fristpage44507
    journal lastpage44507
    identifier eissn1932-619X
    treeJournal of Medical Devices:;2014:;volume( 008 ):;issue: 004
    contenttypeFulltext